FDAFDA · Alimentos y medicamentos·N.º oficial Z-1305-2020
Screw Tap Tapered WP Single-Pat - Product Usage: The Guided Drills (Twist (Step) Drills, Guided Tapered Drills) are used for preparing osteotomies for the specific related implants.
Publicado el 26 de febrero de 2020 · Inicio del retiro: 14 de marzo de 2019
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Due to a problem in the packaging sealing process at the supplier of the affected devices, it cannot be guaranteed that the sterilized medical devices in the scope of this recall remain sterile during their shelf-life.
Qué hacer
Voluntary: Firm initiated · E-Mail
Detalles del retiro
- Fabricante
- Nobel Biocare Usa Llc
- Distribuidor
- Worldwide distribution - US Nationwide distribution and countries of CA, AU, JP, BR, ZA, RU, TW, HK, IN, CN, SA, NZ, MO, CO, LB, AT, BE, CH, CY, DE, DK, EE, ES, FI, FR, GB, GG, GR, HR, HU, IE, LT, LU, NL, PL, PT, RS, SE.
- Lote
- (01)07332747065701(10)73499 Catalog# 37052
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.