VITROS Immunodiagnostic Products CK-MB Calibrators, REF 128 6293, IVD, Ortho-Clinical Diagnostics High Wycombe HP12 4DP, UK For use in the calibration of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System for the quantitative measurement of CK-MB in human serum and plasma (EDTA or heparin).
Publicado el 29 de mayo de 2013 · Inicio del retiro: 3 de abril de 2013
Resumen
FDA registra el retiro Z-1337-2013 de dispositivos médicos de Clase III, publicado el 29 de mayo de 2013 e iniciado el 3 de abril de 2013. Empresa responsable: Ortho-Clinical Diagnostics; Estado: Finalizado; Cantidad reportada: Domestic: 3019 units, Foreign: 905 units; Lote: Lots 1400 (exp. 22 April 2013), 1410 (exp. 20 May 2013), 1420 (exp. 12 July 2013), 1430 (exp. 5 August 2013) , 1440 (exp. 10 September 2013), 1450 (exp. 18 November 2013), 1460 (exp. 18 November 2013), 1465 (exp. 13 November 2013), 1480 (exp. 3 January 2014). La Clase III indica que no es probable que el producto cause consecuencias adversas para la salud.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Internal testing demonstrated that opened-reconstituted VITROS CK-MB Calibrators stored at -20 degrees Celsius (-4 degrees Farenheit) generated invalid calibrations that were detected by positively biased quality control results.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Ortho-Clinical Diagnostics
- Distribuidor
- Worldwide Distribution - Nationwide including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
- Lote
- Lots 1400 (exp. 22 April 2013), 1410 (exp. 20 May 2013), 1420 (exp. 12 July 2013), 1430 (exp. 5 August 2013) , 1440 (exp. 10 September 2013), 1450 (exp. 18 November 2013), 1460 (exp. 18 November 2013), 1465 (exp. 13 November 2013), 1480 (exp. 3 January 2014)
- Clasificación de riesgo
- Class III
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAVITROS Performance Verifier II3 de abril de 2024
- FDAVITROS Performance Verifier I3 de abril de 2024
- FDAVITROS XT Chemistry Products TBIL-ALKP Slides (ALKP is not affected) In Vitro Diagnostic use only - The TBIL2 de septiembre de 2020
- FDAVITROS Chemistry Product Performance Verifier I, Product Code 8231474, UDI # 10758750004577 - Product Usage:24 de junio de 2020
- FDAMyRay RXDC dental unit x-ray unit29 de mayo de 2013
- FDABrilliance CT 16-Slice Water, System Code: 728245; Brilliance CT 10-Slice Water, System Code: 728250;29 de mayo de 2013
Preguntas sobre este retiro
¿VITROS Immunodiagnostic Products CK-MB Calibrators, REF 128 6293, IVD, Ortho-Clinical Diagnostics High Wycombe HP12 4DP, UK For use in the calibration of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System for the quantitative measurement of CK-MB in human serum and plasma (EDTA or heparin). fue retirado del mercado?
Sí. FDA publicó el retiro Z-1337-2013 el 29 de mayo de 2013. Peligro reportado: Internal testing demonstrated that opened-reconstituted VITROS CK-MB Calibrators stored at -20 degrees Celsius (-4 degrees Farenheit) generated invalid calibrations that were detected by positively biased quality control results.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter