VITROS Immunodiagnostic Products CK-MB Calibrators, REF 128 6293, IVD, Ortho-Clinical Diagnostics High Wycombe HP12 4DP, UK For use in the calibration of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System for the quantitative measurement of CK-MB in human serum and plasma (EDTA or heparin).

Publicado el 29 de mayo de 2013 · Inicio del retiro: 3 de abril de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Internal testing demonstrated that opened-reconstituted VITROS CK-MB Calibrators stored at -20 degrees Celsius (-4 degrees Farenheit) generated invalid calibrations that were detected by positively biased quality control results.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Ortho-Clinical Diagnostics
Distribuidor
Worldwide Distribution - Nationwide including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
Lote
Lots 1400 (exp. 22 April 2013), 1410 (exp. 20 May 2013), 1420 (exp. 12 July 2013), 1430 (exp. 5 August 2013) , 1440 (exp. 10 September 2013), 1450 (exp. 18 November 2013), 1460 (exp. 18 November 2013), 1465 (exp. 13 November 2013), 1480 (exp. 3 January 2014)
Clasificación de riesgo
Class III
Área (FDA)
device
País
United States