VITROS Immunodiagnostic Products CK-MB Calibrators, REF 128 6293, IVD, Ortho-Clinical Diagnostics High Wycombe HP12 4DP, UK For use in the calibration of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System for the quantitative measurement of CK-MB in human serum and plasma (EDTA or heparin).

Publicado el 29 de mayo de 2013 · Inicio del retiro: 3 de abril de 2013

Resumen

FDA registra el retiro Z-1337-2013 de dispositivos médicos de Clase III, publicado el 29 de mayo de 2013 e iniciado el 3 de abril de 2013. Empresa responsable: Ortho-Clinical Diagnostics; Estado: Finalizado; Cantidad reportada: Domestic: 3019 units, Foreign: 905 units; Lote: Lots 1400 (exp. 22 April 2013), 1410 (exp. 20 May 2013), 1420 (exp. 12 July 2013), 1430 (exp. 5 August 2013) , 1440 (exp. 10 September 2013), 1450 (exp. 18 November 2013), 1460 (exp. 18 November 2013), 1465 (exp. 13 November 2013), 1480 (exp. 3 January 2014). La Clase III indica que no es probable que el producto cause consecuencias adversas para la salud.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Internal testing demonstrated that opened-reconstituted VITROS CK-MB Calibrators stored at -20 degrees Celsius (-4 degrees Farenheit) generated invalid calibrations that were detected by positively biased quality control results.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Ortho-Clinical Diagnostics
Distribuidor
Worldwide Distribution - Nationwide including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Panama, Singapore, Spain and Venezuela.
Lote
Lots 1400 (exp. 22 April 2013), 1410 (exp. 20 May 2013), 1420 (exp. 12 July 2013), 1430 (exp. 5 August 2013) , 1440 (exp. 10 September 2013), 1450 (exp. 18 November 2013), 1460 (exp. 18 November 2013), 1465 (exp. 13 November 2013), 1480 (exp. 3 January 2014)
Clasificación de riesgo
Class III
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿VITROS Immunodiagnostic Products CK-MB Calibrators, REF 128 6293, IVD, Ortho-Clinical Diagnostics High Wycombe HP12 4DP, UK For use in the calibration of the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System for the quantitative measurement of CK-MB in human serum and plasma (EDTA or heparin). fue retirado del mercado?

Sí. FDA publicó el retiro Z-1337-2013 el 29 de mayo de 2013. Peligro reportado: Internal testing demonstrated that opened-reconstituted VITROS CK-MB Calibrators stored at -20 degrees Celsius (-4 degrees Farenheit) generated invalid calibrations that were detected by positively biased quality control results.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter