INTEGRAL WAND W/TUBING, STERILE, REF 24185, QTY 24/CS, STERILE EO Intended use: The smoke evacuator and accessories are intended to evacuate the smoke created by electrosurgery, laser surgery, or power tool surgical procedures. Locations include Operating Rooms, Trauma, Endoscopy and Laparoscopy Sites.
Publicado el 8 de abril de 2015 · Inicio del retiro: 10 de marzo de 2015
Resumen
FDA registra el retiro Z-1349-2015 de dispositivos médicos de Clase II, publicado el 8 de abril de 2015 e iniciado el 10 de marzo de 2015. Empresa responsable: Microtek Medical Inc; Estado: Finalizado; Cantidad reportada: 9168 units; Lote: Lot Numbers: C10081, C10138, C10181, C10231, C10235, C11269, C11311, C12005, C12051, C12065, C12143, C12172, C12199, C12363, C13015, C13085, C13148, C13189, C13213, C13240, C13295, C13319, C13365, C14104, C14167. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
This recall is being conducted because a cohesive band used to package the devices contains latex and the product is not labeled as required by 21 CFR 801.437; it does not contain Caution: The Packaging of This Product Contains Natural Rubber Latex Which May Cause Allergic Reactions.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Microtek Medical Inc
- Distribuidor
- US (nationwide) Distribution.
- Lote
- Lot Numbers: C10081, C10138, C10181, C10231, C10235, C11269, C11311, C12005, C12051, C12065, C12143, C12172, C12199, C12363, C13015, C13085, C13148, C13189, C13213, C13240, C13295, C13319, C13365, C14104, C14167
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAVACUUM TUBING 7/8" X 6' (2.22cm X 182.88cm), REF 24009, QTY 24\CS, NON-STERILE Intended use: The smoke8 de abril de 2015
- FDAVitagel; indicated in surgical procedures (other than in neurosurgical and ophthalmic) as an adjunct to8 de abril de 2015
- FDASTAGECOACH SALSA CLASSIC SAN ANTONIO SALSAS MISS MICHELE'S PEACH SALSA Net Wt. 16 oz. (454g) Manufactured8 de abril de 2015
- FDASyngo.plaza. A Picture Archiving and Communication System (PACS) intended to display, process, read, report,8 de abril de 2015
- FDANatural Shaped Fried Egg (patties), 1.75 oz. frozen egg patties, packaged 144-count per case, Net WT. 15.758 de abril de 2015
- FDAVarious Lots of Expel Drainage Catheter with TwistLoc Hub Biliary Drainage Catheter and Biliary Drainage8 de abril de 2015
Preguntas sobre este retiro
¿INTEGRAL WAND W/TUBING, STERILE, REF 24185, QTY 24/CS, STERILE EO Intended use: The smoke evacuator and accessories are intended to evacuate the smoke created by electrosurgery, laser surgery, or power tool surgical procedures. Locations include Operating Rooms, Trauma, Endoscopy and Laparoscopy Sites. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1349-2015 el 8 de abril de 2015. Peligro reportado: This recall is being conducted because a cohesive band used to package the devices contains latex and the product is not labeled as required by 21 CFR 801.437; it does not contain Caution: The Packaging of This Product Contains Natural Rubber Latex Which May Cause Allergic Reactions.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter