Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1355-2015

Various Lots of Expel Drainage Catheter with TwistLoc Hub Biliary Drainage Catheter and Biliary Drainage Catheter Kits

Publicado el 8 de abril de 2015 · Inicio del retiro: 25 de febrero de 2015

Resumen

FDA registra el retiro Z-1355-2015 de dispositivos médicos de Clase II, publicado el 8 de abril de 2015 e iniciado el 25 de febrero de 2015. Empresa responsable: Boston Scientific Corporation; Estado: Finalizado; Cantidad reportada: 289; Lote: UPN # LOT # H7493933108200 17245003, 17416605, 17489192 H7493933108350 17245004, 17245005, 17370709, 17408756 H749393310835K1 17411333, 17493554 H7493933110200 17318047 H7493933110350 17275822, 17438003, 17494025, 17501572 H749393311035K1 17328808, 17403871, 17416578, 17459144 H7493933110400 17275824, 17416608 H7493933112350 17245006, 17259580, 17275819 H7493933114350 17275815, 17338207 with Expiration Dates between January 30, 2016 and May 8, 2016. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Complaints of device fragmentation after the catheter was implanted in the biliary system. The most serious patient risk for this issue is additional intervention for fragment retrieval using minimally invasive methods and, in some of the reported cases, this was performed to retrieve the device fragments.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Boston Scientific Corporation
Distribuidor
Worldwide Distribution - US Distribution to the states of : CA, WI, FL and OH., and to the countries of: Belgium, Czech Republic, Finland, France, Germany, Great Britain, Italy, Spain, Switzerland, Austria, Hungary, Netherlands and Portugal.
Lote
UPN # LOT # H7493933108200 17245003, 17416605, 17489192 H7493933108350 17245004, 17245005, 17370709, 17408756 H749393310835K1 17411333, 17493554 H7493933110200 17318047 H7493933110350 17275822, 17438003, 17494025, 17501572 H749393311035K1 17328808, 17403871, 17416578, 17459144 H7493933110400 17275824, 17416608 H7493933112350 17245006, 17259580, 17275819 H7493933114350 17275815, 17338207 with Expiration Dates between January 30, 2016 and May 8, 2016
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Various Lots of Expel Drainage Catheter with TwistLoc Hub Biliary Drainage Catheter and Biliary Drainage Catheter Kits fue retirado del mercado?

Sí. FDA publicó el retiro Z-1355-2015 el 8 de abril de 2015. Peligro reportado: Complaints of device fragmentation after the catheter was implanted in the biliary system. The most serious patient risk for this issue is additional intervention for fragment retrieval using minimally invasive methods and, in some of the reported cases, this was performed to retrieve the device fragments.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter