Protective Cover Vivo 55/65 (Used for Vivo 50/65), Part No. 006344. It is an accessory (i.e. protective cover) for the Vivo 50 and Vivo 65 ventilators - Product Usage: The protective cover is intended for additional protection of the Vivo 50, Vivo 55, Vivo 60 and Vivo 65 ventilators during transportation, and in hospital, institutional, and home care environments. It can be used while the Vivo ventilators is operating, for example mounted on a wheelchair, in a personal vehicle, or carried by hand. The device is not to be used with the protective cover while positioned in a warm place, such as in direct sunlight or close to a radiator. The protective cover includes the following features: transparent window for accessing front panel and buttons; port for patient circuit, cables, O2 inlet, and standby button; cooling air inlet; patient air inlet; straps for safe mounting; and cooling air outlet.
Publicado el 4 de marzo de 2020 · Inicio del retiro: 28 de enero de 2020
Resumen
FDA registra el retiro Z-1352-2020 de dispositivos médicos de Clase II, publicado el 4 de marzo de 2020 e iniciado el 28 de enero de 2020. Empresa responsable: Breas Medical, Inc.; Estado: Finalizado; Cantidad reportada: 629; Lote: The cover does not have a lot number; these covers were distributed between 01/11/2018 and 12/17/2019. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The outlet hole of the exhalation valve control pressure tube on the ventilator may become obstructed by the protective cover.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Breas Medical, Inc.
- Distribuidor
- Worldwide distribution - US Nationwide distribution in the states of AR, CA, FL, GA, IA, IL, KS, MA, MI, MO, NJ, NY, OH, PR, SC, TX, VA, WI and countries distribution worldwide.
- Lote
- The cover does not have a lot number; these covers were distributed between 01/11/2018 and 12/17/2019
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAVivo 45 LS, pressure and volume ventilator capable of delivering continuous or intermittent ventilatory11 de septiembre de 2024
- FDAEllipse, Tiered-therapy Cardioverter/Defibrillator, REF (UDI/GTIN): CD1377-36QC (05414734507646), CD2411-36C4 de marzo de 2020
- FDAZimmer Sterile Disposable Tourniquet Cuff with Protective Sleeve and PLC, Dual Port, Dual Bladder, Item No.4 de marzo de 2020
- FDATHP Hip Plating System, Item Nos. 14-451070 14-451075 14-451080 14-451085 14-451090 14-451095 14-4511004 de marzo de 2020
- FDAMedline ROBOTIC-LF REF CDS984543B Lot 20AKA927 Expiration 2020-09-30 GTIN (01)108882770563504 de marzo de 2020
- FDAPhilips DigiTrak XT Holter Recorder Software version number B.1 and C.2. The DigiTrak XT Recorder is available4 de marzo de 2020
Preguntas sobre este retiro
¿Protective Cover Vivo 55/65 (Used for Vivo 50/65), Part No. 006344. It is an accessory (i.e. protective cover) for the Vivo 50 and Vivo 65 ventilators - Product Usage: The protective cover is intended for additional protection of the Vivo 50, Vivo 55, Vivo 60 and Vivo 65 ventilators during transportation, and in hospital, institutional, and home care environments. It can be used while the Vivo ventilators is operating, for example mounted on a wheelchair, in a personal vehicle, or carried by hand. The device is not to be used with the protective cover while positioned in a warm place, such as in direct sunlight or close to a radiator. The protective cover includes the following features: transparent window for accessing front panel and buttons; port for patient circuit, cables, O2 inlet, and standby button; cooling air inlet; patient air inlet; straps for safe mounting; and cooling air outlet. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1352-2020 el 4 de marzo de 2020. Peligro reportado: The outlet hole of the exhalation valve control pressure tube on the ventilator may become obstructed by the protective cover.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter