Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed.

Publicado el 5 de junio de 2013 · Inicio del retiro: 30 de abril de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Elekta, Inc.
Distribuidor
USA Nationwide including the states of CA, FL, IL, MD, MA, NJ, NY, NC, ND, OH, OR, PA, TN, TX, VA, WA, DC and Puerto Rico
Lote
325134, 325254, 325082, 325159, 325268, 325237, 325331, 325267, 325236, 325314, 325127, 325261, 325286, 325077, 325269, 325129, 325296, 325273, 325139, 325201, 325137, 325184, 325161, 325076, 325086, 325112, 325081, 325083, 325099, 325110, 325271
Clasificación de riesgo
Class II