Leucadia Autolok" Polyaxial Pedicle Screw System. Intended to help provide correction, immobilization and stabilization of spinal segments.

Publicado el 5 de junio de 2013 · Inicio del retiro: 6 de mayo de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Alphatec Spine initiated this medical device recall for the following reason: due to tolerancing, the flange of the Phygen Leucadia" Autolok set screw may interfere with the top of the screw head prior to fully locking the rod. If this condition occurs, the polyaxial screw head and rod may fail to provide a rigid construct under loading. To date there have been no reports of patient injury, adve

Qué hacer

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Detalles del retiro

Fabricante
Alphatec Spine, Inc.
Distribuidor
Nationwide Distribution.
Lote
83010 85001 85002 83005-30 83005-35 83005-40 83005-45 83005-50 83006-30 83006-35 83006-40 83006-45 83006-50 83006-55 83006-60 83007-30 83007-35 83007-40 83007-45 83007-50 83007-55 83008-30 83008-35 83008-40 83008-45 83008-50 83008-55 83008-80 82001-1 Catalog Numbers: LA-HSS0010 LA-HDOO01 LA-CTOO02 LA-PS0530 LA-PS0535 LA-PS0540 LA-PS0545 LA-PS0550 LA-PS0630 LA-PS0635 LA-PS0640 LA-PS0645 LA-PS0650 LA-PS0655 LA-PS0660 LA-PS0730 LA-PS0735 LA-PS0740 LA-PS0745 LA-PS0750 LA-PS0755 LA-PS0830 LA-PS0835 LA-PS0840 LA-PS0845 LA-PS0850 LA-PS0855 LA-PS0880 LA-SOO01 Lots Affected: All
Clasificación de riesgo