FDAFDA · Alimentos y medicamentos·N.º oficial Z-1425-2022

Sensis with software version VD12A. A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology Model:10764561

Publicado el 27 de julio de 2022 · Inicio del retiro: 15 de junio de 2022

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Sensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivity to the Sensis Vibe system resulting in no vital signs being available.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Siemens Medical Solutions USA, Inc
Distribuidor
US Nationwide Distribution AZ FL KY MI OH PR TX
Lote
UDI-DI: 04056869010137 Serial Numbers: 61331, 60015
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States