Soluble Transferrin Receptors (STFR), Catalog Number TF10159 (Research Use Only). IVD
Publicado el 27 de julio de 2022 · Inicio del retiro: 9 de junio de 2022
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in non-numerical results. This affects Siemens Atellica CH analysers only. Customer have reported an atypical calibration curve that results in the values around the low QC range to produce a non-numerical result. It appears to be related to the bottle values of the calibrator producing a curve shape that is too linear for the logitlog calculation to calculate. This issue can only be identified in the Atellica graphical view of the calibration. In most of the cases when this has occurred, it is picked up by the QC generating a Math Error, however, there are reports of QC shifts after calibration without the Math Error flagged that may be related to this issue. Results reported with STFR Level 1, with a concentration less than 1.7mg/L, using TF10159 and TF10161 can be underestimated by up to 30%. This issue has been reported with reagent lots 575475 and 562691, calibrator lots 537725 and 562696. No other complaints received.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Randox Laboratories Ltd.
- Distribuidor
- US Nationwide distribution in the state of IL.
- Lote
- UDI-DI (GTIN): 05055273215564 All lots within expiry
- Clasificación de riesgo
- Class III
- Área (FDA)
- device
- País
- Ireland