Alinity i Processing Module, 03R65 01, with the below components. a. TED (Thermo Electric Device) Engine, Reagent cooler; b. CABLE, POWER OUT, TED CNTRLER Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.

Publicado el 5 de junio de 2019 · Inicio del retiro: 5 de abril de 2019

Resumen

FDA registra el retiro Z-1497-2019 de dispositivos médicos de Clase II, publicado el 5 de junio de 2019 e iniciado el 5 de abril de 2019. Empresa responsable: Abbott Gmbh & Co. KG; Estado: Finalizado; Cantidad reportada: 373 units total; Lote: Part Number: 03R65 01, all Serial Numbers; a. TED (Thermo Electric Device) Engine, Reagent cooler - Part Numbers A 30103732 01, A 30103732 02 with all Serial Numbers; b. CABLE, POWER OUT, TED CNTRLER - Part Numbers A 35006203 01, A 35006203 02 with all Serial Numbers. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Potential loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. Could cause burn and shock injuries to users and a delay in test results. Device design has mitigated this risk, but it is a possibility.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Abbott Gmbh & Co. KG
Distribuidor
Worldwide Distribution - US Nationwide to Oklahoma and Texas. Distributed worldwide to ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, MADAGASCAR, MEXICO, NETHERLANDS, NORWAY, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SERBIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UAE, UNITED KINGDOM, VIETNAM.
Lote
Part Number: 03R65 01, all Serial Numbers; a. TED (Thermo Electric Device) Engine, Reagent cooler - Part Numbers A 30103732 01, A 30103732 02 with all Serial Numbers; b. CABLE, POWER OUT, TED CNTRLER - Part Numbers A 35006203 01, A 35006203 02 with all Serial Numbers
Clasificación de riesgo
Class II
Área (FDA)
device
País
Germany

Retiros relacionados

Preguntas sobre este retiro

¿Alinity i Processing Module, 03R65 01, with the below components. a. TED (Thermo Electric Device) Engine, Reagent cooler; b. CABLE, POWER OUT, TED CNTRLER Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1497-2019 el 5 de junio de 2019. Peligro reportado: Potential loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. Could cause burn and shock injuries to users and a delay in test results. Device design has mitigated this risk, but it is a possibility.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter