Alinity i Processing Module, 03R65 01, with the below components. a. TED (Thermo Electric Device) Engine, Reagent cooler; b. CABLE, POWER OUT, TED CNTRLER Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.
Publicado el 5 de junio de 2019 · Inicio del retiro: 5 de abril de 2019
Resumen
FDA registra el retiro Z-1497-2019 de dispositivos médicos de Clase II, publicado el 5 de junio de 2019 e iniciado el 5 de abril de 2019. Empresa responsable: Abbott Gmbh & Co. KG; Estado: Finalizado; Cantidad reportada: 373 units total; Lote: Part Number: 03R65 01, all Serial Numbers; a. TED (Thermo Electric Device) Engine, Reagent cooler - Part Numbers A 30103732 01, A 30103732 02 with all Serial Numbers; b. CABLE, POWER OUT, TED CNTRLER - Part Numbers A 35006203 01, A 35006203 02 with all Serial Numbers. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Potential loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. Could cause burn and shock injuries to users and a delay in test results. Device design has mitigated this risk, but it is a possibility.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Abbott Gmbh & Co. KG
- Distribuidor
- Worldwide Distribution - US Nationwide to Oklahoma and Texas. Distributed worldwide to ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, INDONESIA, ISRAEL, ITALY, JAPAN, MADAGASCAR, MEXICO, NETHERLANDS, NORWAY, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SERBIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TURKEY, UAE, UNITED KINGDOM, VIETNAM.
- Lote
- Part Number: 03R65 01, all Serial Numbers; a. TED (Thermo Electric Device) Engine, Reagent cooler - Part Numbers A 30103732 01, A 30103732 02 with all Serial Numbers; b. CABLE, POWER OUT, TED CNTRLER - Part Numbers A 35006203 01, A 35006203 02 with all Serial Numbers
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Germany
Retiros relacionados
- FDALN 3R70-01; software version 2.6.2 and earlier. The software is used with the Alinity i (LN 03R65-01)1 de abril de 2020
- FDAARCHITECT HAVAB-G Calibrator, REF 6L27-014 de diciembre de 2019
- FDASOLOPATH Balloon Expandable TransFemoral System, Model Numbers STFI-1425, STFI-1435, STFI-1625, STFI-1635,5 de junio de 2019
- FDAMaquet Getinge-BEQ-TOP 8013 ECC QUADROX iD Pack RF32 Material: 7010479625 de junio de 2019
- FDALaxmi Brand Dried Apricot 7 oz (200g) Plastic Pouch UPC Code: 23246170545 de junio de 2019
- FDAAbbott Alinity ci-series System Control Module software version 2.5.1; LN 3R70-01 with the following hardware:22 de mayo de 2019
Preguntas sobre este retiro
¿Alinity i Processing Module, 03R65 01, with the below components. a. TED (Thermo Electric Device) Engine, Reagent cooler; b. CABLE, POWER OUT, TED CNTRLER Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1497-2019 el 5 de junio de 2019. Peligro reportado: Potential loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. Could cause burn and shock injuries to users and a delay in test results. Device design has mitigated this risk, but it is a possibility.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter