Echelon, Echelon Oval and Oasis MR imaging device systems; one unit per box. The Echelon, Echelon Oval and Oasis MR systems are imaging devices, and are intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR systems produce transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities.
Publicado el 26 de junio de 2013 · Inicio del retiro: 31 de enero de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Hitachi discovered a software error that can occur when simultaneously scanning a patient while processing images previously acquired from a different patient. The software error can occur randomly. The error causes an image in the processed dataset to have the same unique identifier (UID) as in the scanning dataset. The processed image is stored in the place of the original image, which is lost.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Hitachi Medical Systems America Inc
- Distribuidor
- Worldwide Distribution - US Nationwide including the states of AZ, CA, CO, DE, FL, IL, IN, KS, KY, MD, NC, ND, NY, OH, OR, PA, TX, VA, WA, and WI. One (1) system unit was also distributed to Mexico.
- Lote
- The following serial numbers are associated with this recall: Echelon - V061, V062, and V065 Oasis - M009, M017, M067, M071, M084, M106-M108, M110-M141 inclusive, and M952 Echelon Oval - Y001-Y005 inclusive
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.