Echelon, Echelon Oval and Oasis MR imaging device systems; one unit per box. The Echelon, Echelon Oval and Oasis MR systems are imaging devices, and are intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR systems produce transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities.
Publicado el 26 de junio de 2013 · Inicio del retiro: 31 de enero de 2013
Resumen
FDA registra el retiro Z-1559-2013 de dispositivos médicos de Clase II, publicado el 26 de junio de 2013 e iniciado el 31 de enero de 2013. Empresa responsable: Hitachi Medical Systems America Inc; Estado: Finalizado; Cantidad reportada: 45 system units; Lote: The following serial numbers are associated with this recall: Echelon - V061, V062, and V065 Oasis - M009, M017, M067, M071, M084, M106-M108, M110-M141 inclusive, and M952 Echelon Oval - Y001-Y005 inclusive. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Hitachi discovered a software error that can occur when simultaneously scanning a patient while processing images previously acquired from a different patient. The software error can occur randomly. The error causes an image in the processed dataset to have the same unique identifier (UID) as in the scanning dataset. The processed image is stored in the place of the original image, which is lost.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Hitachi Medical Systems America Inc
- Distribuidor
- Worldwide Distribution - US Nationwide including the states of AZ, CA, CO, DE, FL, IL, IN, KS, KY, MD, NC, ND, NY, OH, OR, PA, TX, VA, WA, and WI. One (1) system unit was also distributed to Mexico.
- Lote
- The following serial numbers are associated with this recall: Echelon - V061, V062, and V065 Oasis - M009, M017, M067, M071, M084, M106-M108, M110-M141 inclusive, and M952 Echelon Oval - Y001-Y005 inclusive
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAHitachi Echelon MRI System29 de marzo de 2017
- FDAHitachi Oasis MRI System29 de marzo de 2017
- FDAHitachi Oasis MRI System - C-Spine Coil1 de febrero de 2017
- FDAHitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended to10 de septiembre de 2014
- FDAGlideScope Video Laryngoscope (GVL) 3, 4, 5 and GlideScope (AVL) 2, 3, 4 and 5 reusable blades Product Usage:26 de junio de 2013
- FDATecan Freedom EVO with Touch Tools Suite Version 3.0 The intended use of the Touch Tools Suite is to run26 de junio de 2013
Preguntas sobre este retiro
¿Echelon, Echelon Oval and Oasis MR imaging device systems; one unit per box. The Echelon, Echelon Oval and Oasis MR systems are imaging devices, and are intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR systems produce transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1559-2013 el 26 de junio de 2013. Peligro reportado: Hitachi discovered a software error that can occur when simultaneously scanning a patient while processing images previously acquired from a different patient. The software error can occur randomly. The error causes an image in the processed dataset to have the same unique identifier (UID) as in the scanning dataset. The processed image is stored in the place of the original image, which is lost.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter