Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1558-2013

GlideScope Video Laryngoscope (GVL) 3, 4, 5 and GlideScope (AVL) 2, 3, 4 and 5 reusable blades Product Usage: The GlideScope Video Laryngoscopes are intended for use by qualified medical professionals to obtain a clear, unobstructed view of the vocal cords for medical procedures. The GlideScope Video Laryngoscopes system should only be used by individuals who have been trained and authorized by a physician or the institution providing patient care. The device is not life sustaining or life supporting. The GlideScope AVL Video Monitor system has the same intended use as the GVL System. The difference between systems is that the AVL has a digital platform and the GVL an analog platform.

Publicado el 26 de junio de 2013 · Inicio del retiro: 8 de mayo de 2013

Resumen

FDA registra el retiro Z-1558-2013 de dispositivos médicos de Clase I, publicado el 26 de junio de 2013 e iniciado el 8 de mayo de 2013. Empresa responsable: Verathon, Inc.; Estado: Finalizado; Cantidad reportada: 4784 units worldwide (3,381 in the US and 1403 international); Lote: Model, Part and Serial Numbers: Model # GVL 3, Part # 0574-0007, Serial # range MD112388 to MD121908; Model # GVL 4 Part # 0574-0001 Serial # range LG112759 to LG122582; Model # GVL 5 Part # 0574-0030 Serial # range XL111799 to XL121759; Model # AVL 2 Part # 0574-0118 Serial # range AC111500 to AC121604; Model # AVL 3 Part # 0574-0115 Serial # range AD111500 to AD121688; Model # AVL 4 Part # 0574-0116 Serial # range AE11150 to AE121778; and Model # AVL 5 Part # 0574-0117 Serial # range AF111500 to AF121666. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Verathon GlideScope Reusable GVL and AVL Blades are being recalled due to the potential risk of premature failure/breakage of the blade tip that may not be readily visible during routine inspection before or after intubation.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Verathon, Inc.
Distribuidor
Worldwide distribution: USA (nationwide) including states of : AK, AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Washington DC, Guam, and countries of : Argentina, Australia, Brazil, Canada, China, Ecuador, France, Germany, Hong Kong, India, Indonesia, Italy, Japan, Japan, Malaysia, Panama, Peru, Philippines, Singapore, South Korea, Taiwan, Thailand, The Netherlands, United Kingdom, and Venezuela.
Lote
Model, Part and Serial Numbers: Model # GVL 3, Part # 0574-0007, Serial # range MD112388 to MD121908; Model # GVL 4 Part # 0574-0001 Serial # range LG112759 to LG122582; Model # GVL 5 Part # 0574-0030 Serial # range XL111799 to XL121759; Model # AVL 2 Part # 0574-0118 Serial # range AC111500 to AC121604; Model # AVL 3 Part # 0574-0115 Serial # range AD111500 to AD121688; Model # AVL 4 Part # 0574-0116 Serial # range AE11150 to AE121778; and Model # AVL 5 Part # 0574-0117 Serial # range AF111500 to AF121666.
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿GlideScope Video Laryngoscope (GVL) 3, 4, 5 and GlideScope (AVL) 2, 3, 4 and 5 reusable blades Product Usage: The GlideScope Video Laryngoscopes are intended for use by qualified medical professionals to obtain a clear, unobstructed view of the vocal cords for medical procedures. The GlideScope Video Laryngoscopes system should only be used by individuals who have been trained and authorized by a physician or the institution providing patient care. The device is not life sustaining or life supporting. The GlideScope AVL Video Monitor system has the same intended use as the GVL System. The difference between systems is that the AVL has a digital platform and the GVL an analog platform. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1558-2013 el 26 de junio de 2013. Peligro reportado: Verathon GlideScope Reusable GVL and AVL Blades are being recalled due to the potential risk of premature failure/breakage of the blade tip that may not be readily visible during routine inspection before or after intubation.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter