Siemens ADVIA Centaur Folate Cat No. 06891541 SMN 10325366 (500 tests) Cat No. 06367974 SMN 10310308 (100 tests) Cat No. 00203473 SMN 10331250 (Ref) (500 tests) Cat No. 09132781 SMN 10340209 (Ref) (2500 tests) For in vitro diagnostic use in the quantitative determination of folate in serum or red blood cells using the ADVIA Centaur and ADVIA Centaur XP systems
Publicado el 14 de mayo de 2014 · Inicio del retiro: 25 de marzo de 2014
Resumen
FDA registra el retiro Z-1564-2014 de dispositivos médicos de Clase II, publicado el 14 de mayo de 2014 e iniciado el 25 de marzo de 2014. Empresa responsable: Siemens Healthcare Diagnostics, Inc; Estado: Finalizado; Cantidad reportada: 49123 units; Lote: Kit Lots ending in and exp date: 219 Exp. Date 25 March, 2014; 222 Exp Date 12 May, 2014; 224 Exp Date 04 August, 2014; 225 Exp Date 04 August, 2014; 226 Exp Date 12 September, 2014; 227 Exp. Date 06 October, 2014. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Serum controls may go out of range low with the ADVIA Centaur ¿Systems Folate assay and a shift in whole blood patient may result with kit lots ending in 219 and 223
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Siemens Healthcare Diagnostics, Inc
- Distribuidor
- Worldwide Distribution-USA (nationwide) and the countries of Austria Belgium Bulgaria Cyprus Czech Republic Denmark Estonia Finland Finland France Fren.Polynesia Germany Great Britain Greece Guadeloupe Hungary Iceland Ireland Israel Italy Latvia Lithuania Luxembourg Malta Martinique Netherlands Norway Poland Portugal Qatar Reunion Romania Russian Fed. Serbia Slovakia Slovenia Spain Sweden Switzerland Turkey Unit.Arab Emir.
- Lote
- Kit Lots ending in and exp date: 219 Exp. Date 25 March, 2014; 222 Exp Date 12 May, 2014; 224 Exp Date 04 August, 2014; 225 Exp Date 04 August, 2014; 226 Exp Date 12 September, 2014; 227 Exp. Date 06 October, 2014
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDASiemens ADVIA Centaur Folate 100 test kit-for IVD of folate in serum or red blood cells SMN:1031030829 de enero de 2020
- FDASiemensADVIA Centaur aTG (500 test kit) Siemens Material Number (SMN): 104923998 de mayo de 2019
- FDAAtellica IM aTG (500 test kit) Siemens Material Number (SMN): 109954628 de mayo de 2019
- FDASiemens Atellica IM aTG (100 test kit) Siemens Material Number (SMN): 109954618 de mayo de 2019
- FDAMar Cor Purification, Central Water Platform (CWP), 102, 104 and 106. Product Usage: The CWP is a Central14 de mayo de 2014
- FDASpineFrontier S-Lift Lock Shaft Inserter, an instrument used with S-LIFT Intervertebral Body Fusion Device in14 de mayo de 2014
Preguntas sobre este retiro
¿Siemens ADVIA Centaur Folate Cat No. 06891541 SMN 10325366 (500 tests) Cat No. 06367974 SMN 10310308 (100 tests) Cat No. 00203473 SMN 10331250 (Ref) (500 tests) Cat No. 09132781 SMN 10340209 (Ref) (2500 tests) For in vitro diagnostic use in the quantitative determination of folate in serum or red blood cells using the ADVIA Centaur and ADVIA Centaur XP systems fue retirado del mercado?
Sí. FDA publicó el retiro Z-1564-2014 el 14 de mayo de 2014. Peligro reportado: Serum controls may go out of range low with the ADVIA Centaur ¿Systems Folate assay and a shift in whole blood patient may result with kit lots ending in 219 and 223
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter