Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1615-2024

LightStrike Blackout Curtains, Part numbers; 610-00008, 610-00009, 610-00019, 610-00020, and 610-00024, used with the LightStrike Germ-Zapping Robots Models: PXUV4D, PXUV5D, LSMUV6-FT, or LSMUV6-SL, for whole room microbial reduction.

Publicado el 1 de mayo de 2024 · Inicio del retiro: 18 de marzo de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Blackout curtains are no longer recommended as a protective barrier against UV/Ozone exposure to a bystander standing in the same room, but on the other side of the curtain from where a microbial reduction device UV cycle is taking place. Bystanders in the room during the cycle may experience temporary skin reddening, feeling of sand in eyes, nausea, headache, and/or mucous membrane irritation.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
XENEX Disinfection Services Inc.
Distribuidor
US Nationwide distribution in the states of AR, AZ, CA, CO, FL, GA, HI, ID, IL, KY, LA, MI, MN, MO, NC, ND, NV, OH, OK, OR, TN, UT, WA, WV, WY.
Lote
Robot UDI-DIs: 00814122020033, 00814122020064, 00814122020125, 00814122020132. All serial numbers.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States