ORTHO® Assay Software (OAS) Server, Catalog # 6842840 Usage: The ORTHO® Summit System (OSS) is a modular, integrated system for use with licensed blood screening tests. Consisting of networked processors and instruments linked together via a local area network, assay-specific pipetting protocols and ORTHO Assay Software (OAS), OSS automates many of the processing functions and data management requirements associated with ELISA (enzyme-linked immunosorbent assay) microplate testing.
Publicado el 10 de julio de 2013 · Inicio del retiro: 6 de mayo de 2013
Resumen
FDA registra el retiro Z-1651-2013 de dispositivos médicos de Clase II, publicado el 10 de julio de 2013 e iniciado el 6 de mayo de 2013. Empresa responsable: Ortho-Clinical Diagnostics; Estado: Finalizado; Cantidad reportada: 31 units; Lote: Serial numbers: 0811F06216 through 0811F06225 and 0911F06240 through 0911F06245 and 1011F06317 through 1011F06331. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Ortho-Clinical Diagnostics
- Distribuidor
- Worldwide Distribution - USA Nationwide including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI & WV, and countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, PR, Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain
- Lote
- Serial numbers: 0811F06216 through 0811F06225 and 0911F06240 through 0911F06245 and 1011F06317 through 1011F06331
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAVITROS Performance Verifier II3 de abril de 2024
- FDAVITROS Performance Verifier I3 de abril de 2024
- FDAVITROS XT Chemistry Products TBIL-ALKP Slides (ALKP is not affected) In Vitro Diagnostic use only - The TBIL2 de septiembre de 2020
- FDAVITROS Chemistry Product Performance Verifier I, Product Code 8231474, UDI # 10758750004577 - Product Usage:24 de junio de 2020
- FDAVITROS 3600 Immunodiagnostic System (product code 6802783) Product Usage: For use in the in vitro10 de julio de 2013
- FDALithium Thionyl non-rechargeable disposable battery pack P/N 350-0294-01 used with ER920 Wireless Series10 de julio de 2013
Preguntas sobre este retiro
¿ORTHO® Assay Software (OAS) Server, Catalog # 6842840 Usage: The ORTHO® Summit System (OSS) is a modular, integrated system for use with licensed blood screening tests. Consisting of networked processors and instruments linked together via a local area network, assay-specific pipetting protocols and ORTHO Assay Software (OAS), OSS automates many of the processing functions and data management requirements associated with ELISA (enzyme-linked immunosorbent assay) microplate testing. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1651-2013 el 10 de julio de 2013. Peligro reportado: There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter