Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1671-2020

ARCHITECT C System Mixer - Product Usage: The ARCHITECT Clinical Chemistry System is designed to perform automated chemistry tests, utilizing photometry and potentiometric technology. The mixer is a component and it is used on all ARCHITECT Clinical Chemistry systems to mix the sample with the reagent inside of the cuvettes.

Publicado el 15 de abril de 2020 · Inicio del retiro: 1 de mayo de 2019

Resumen

FDA registra el retiro Z-1671-2020 de dispositivos médicos de Clase II, publicado el 15 de abril de 2020 e iniciado el 1 de mayo de 2019. Empresa responsable: Abbott Laboratories, Inc; Estado: Finalizado; Cantidad reportada: 09D59-01, 1010 products, 09D59-02, 25156 products, 09D59-03 13976 products; Lote: LN 09D59-01, LN 09D59-02, LN 09D59-03. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Mixer blade may separate from mixer due to the screw and nut failure and detachment. As a result of the mixer blade missing, the sample and reagent are inadequately mixed.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Abbott Laboratories, Inc
Distribuidor
Worldwide distribution  US Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, ME, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, Puerto Rico and St. Croix US Virgin Islands. Countries of AF, AL, AO, AR, AM, AU, AT, AZ, BS, BD, BV, BE, BA, BW, BR, BN, BG, BF, CA, CL, CN, CO, CR, HR, CY, CZ, CD, DK, DO, EC, EG, SV, EE, ET, FI, FR, GE, DE, GR, GT, HN, HK, HU, IN, ID, IQ, IE, IL, IT, JM, JO, KZ, KE, KW, KG, LV, LB, LY, LT, LU, MK, MW, MY, MV, ML, MR, MX, MD, MA, MM, NA, NL, NZ, NG, NO, OM, PK, PA, PE, PH, PL, PT, PY, QA, RO, RU, RW, SA, SN, SG, SK, SI, ZA, ES, SE, CH, TW, TJ, TZ, TH, TT, TN, TC, TR, AE, UG, UK, UA, UY, VE, VN, YE, ZM, ZW, LC, MS, ME, VC, LK, RS, KY, CW, GZA, JRH, CI, KR.
Lote
LN 09D59-01, LN 09D59-02, LN 09D59-03
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿ARCHITECT C System Mixer - Product Usage: The ARCHITECT Clinical Chemistry System is designed to perform automated chemistry tests, utilizing photometry and potentiometric technology. The mixer is a component and it is used on all ARCHITECT Clinical Chemistry systems to mix the sample with the reagent inside of the cuvettes. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1671-2020 el 15 de abril de 2020. Peligro reportado: Mixer blade may separate from mixer due to the screw and nut failure and detachment. As a result of the mixer blade missing, the sample and reagent are inadequately mixed.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter