AGFA Digital Radiography X-Ray System DX-D100 DX-D100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis
Publicado el 5 de abril de 2017 · Inicio del retiro: 20 de enero de 2016
Resumen
FDA registra el retiro Z-1698-2017 de dispositivos médicos de Clase II, publicado el 5 de abril de 2017 e iniciado el 20 de enero de 2016. Empresa responsable: AGFA Healthcare Corp.; Estado: Finalizado; Cantidad reportada: 308; Lote: A5410000053, A5410000054, A5411000439, A5411000446, A5411000447, A5411000448, A5410000067, A5410000068, A5410000069, A5410000070, A5410000072, A5410000073, A5410000100, A5410000107, A5410000134, A5411000184, A5410000039, A5411000140, A5411000142, A5411000314, A5411000342, A5411000347, A5411000349, A5410000007, A5411000017, A5411000315, A5411000331, A5411000332, A5411000463, A5411000141, A5411000144, A5411000539, A5411000541, A5411000069, A5411000133, A5411000175, A5411000119, A5411000123, A5411000019, A5411000400, A5411000401, A5411000540, A5411000047, A5411000048, A5411000053, A5410000041, A5410000042, A5410000046, A5411000041, A5411000079, A5411000024, A5411000032, A5411000083, A5411000038, A5411000132, A5411000571, A5411000573, A5411000050, A5411000088, A5411000247, A5411000070, A5411000078, A5411000082, A5411000476, A5411000478, A5411000480, A5411000482, A5411000489, A5411000490, A5411000563, A5411000582, A5411000154, A5411000156, A5411000404, A5411000500, A5411000501, A5411000589, A5411000591, A5411000583, A5411000236, A5411000155, A5411000151, A5411000191, A5411000192, A5411000194, A5411000193, A5411000249, A5411000052, A5411000185, A5411000237, A5411000030, A5411000397, A5411000587, A5411000499, A5411000503, A5411000504, A5411000574, A5411000584, A5411000348, A5411000193, A5411000268, A5411000498, A5411000502, A5411000509, A5411000403, A5411000153, A5411000233, A5411000235, A5411000318, A5411000398, A5411000538, A5411000371, A5410000043, A5411000181, A5411000158, A5411000513, A5411000514, A5411000152, A5411000548, A5411000553, A5411000199, A5411000209, A5411000210, A5411000232, A5411000346, A5411000124, A5411000174, A5411000177, A5411000230, A5411000468, A5411000469, A5411000470, A5411000173, A5411000316, A5411000145, A5411000188, A5411000197, A5411000202, A5411000203, A5411000206, A5411000208, A5411000335, A5411000336, A5411000533, A5411000537, A5411000536, A5411000159, A5410000038, A5411000148, A5411000149, A5411000134, A5411000405, A5411000120, A5411000125, A5411000372, A5411000147, A5411000588, A5411000039, A5411000117, A5411000121, A5411000333, A5411000334, A5411000377, A5411000040, A5411000240, A5411000593, A5411000594, A5411000085, A5411000419, A5411000097, A5411000281, A5411000278, A5411000279, A5411000280, A5411000492, A5411000282, A5411000245, A5411000277, A5411000064, A5411000065, A5411000187, A5411000198, A5411000207, A5411000211, A5411000212, A5411000231, A5411000399, A5411000491, A5411000493. A5411000524, A5411000497, A5411000534, A5411000526, A5411000535, A5411000122, A5411000575, A5411000576, A5411000577, A5411000495, A5411000578, A5410000044, A5411000037, A5411000241, A5411000246, A5411000108, A5411000110, A5411000114, A5411000312, A5411000074, A5411000080, A5411000522, A5411000033, A5411000139, A5411000543, A5411000182, A5411000190, A5411000021, A5411000374, A5411000218, A5411000572, A5411000084, A5411000098, A5411000238, A5411000116, A5411000408, A5411000409, A5410000042, A5410000023, A5411000544, A5411000580, A5411000581, A5411000545, A5411000546, A5411000547, A5411000579, A5411000337, A5411000338, A5411000340, A5411000341, A5411000475, A5411000549, A5411000373, A5411000420, A5410000133, A5410000117, A5410000115, A5410000079, A5410000131, A5410000066, A5410000132, A5410000114, A5410000082, A5410000116, A5411000025, A5411000523, A5411000525, A5411000590, A5411000592, A5411000263, A5411000309, A5411000262, A5411000270, A5411000272, A5411000273, A5411000293, A5411000363, A5411000260, A5411000264, A5411000269, A5411000271, A5411000291, A5411000292, A5411000261, A5411000265, A5411000266, A5411000267, A5411000034, A5411000081, A5411000146, A5411000092, A5411000093, A5411000407, A5411000026, A5411000068, A5411000367, A5411000368, A5411000465, A5411000060, A5411000131, A5411000018, A5411000100, A5411000515, A5411000313, A5411000284, A5411000076, A5411000135, A5411000180, A5411000343, A5411000285, A5411000311, A5411000375, A5411000062, A5411000136, A5411000179, A5411000178, A5411000421, A5411000427, A5411000. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The DX-D100 User Manual already contained information about to move a DX-D100 mobile unit manually by using an Allen wrench, but did not indicate where the Allen wrench should be stored.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- AGFA Healthcare Corp.
- Distribuidor
- Worldwide Distribution - US including AL, AR, CA, CO, DC, FL, GA, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NJ, NY, NC, OH, OK, OR, RI, SC, TN, TX, VA, WV, WI, WY, and Internationally to Canada.
- Lote
- A5410000053, A5410000054, A5411000439, A5411000446, A5411000447, A5411000448, A5410000067, A5410000068, A5410000069, A5410000070, A5410000072, A5410000073, A5410000100, A5410000107, A5410000134, A5411000184, A5410000039, A5411000140, A5411000142, A5411000314, A5411000342, A5411000347, A5411000349, A5410000007, A5411000017, A5411000315, A5411000331, A5411000332, A5411000463, A5411000141, A5411000144, A5411000539, A5411000541, A5411000069, A5411000133, A5411000175, A5411000119, A5411000123, A5411000019, A5411000400, A5411000401, A5411000540, A5411000047, A5411000048, A5411000053, A5410000041, A5410000042, A5410000046, A5411000041, A5411000079, A5411000024, A5411000032, A5411000083, A5411000038, A5411000132, A5411000571, A5411000573, A5411000050, A5411000088, A5411000247, A5411000070, A5411000078, A5411000082, A5411000476, A5411000478, A5411000480, A5411000482, A5411000489, A5411000490, A5411000563, A5411000582, A5411000154, A5411000156, A5411000404, A5411000500, A5411000501, A5411000589, A5411000591, A5411000583, A5411000236, A5411000155, A5411000151, A5411000191, A5411000192, A5411000194, A5411000193, A5411000249, A5411000052, A5411000185, A5411000237, A5411000030, A5411000397, A5411000587, A5411000499, A5411000503, A5411000504, A5411000574, A5411000584, A5411000348, A5411000193, A5411000268, A5411000498, A5411000502, A5411000509, A5411000403, A5411000153, A5411000233, A5411000235, A5411000318, A5411000398, A5411000538, A5411000371, A5410000043, A5411000181, A5411000158, A5411000513, A5411000514, A5411000152, A5411000548, A5411000553, A5411000199, A5411000209, A5411000210, A5411000232, A5411000346, A5411000124, A5411000174, A5411000177, A5411000230, A5411000468, A5411000469, A5411000470, A5411000173, A5411000316, A5411000145, A5411000188, A5411000197, A5411000202, A5411000203, A5411000206, A5411000208, A5411000335, A5411000336, A5411000533, A5411000537, A5411000536, A5411000159, A5410000038, A5411000148, A5411000149, A5411000134, A5411000405, A5411000120, A5411000125, A5411000372, A5411000147, A5411000588, A5411000039, A5411000117, A5411000121, A5411000333, A5411000334, A5411000377, A5411000040, A5411000240, A5411000593, A5411000594, A5411000085, A5411000419, A5411000097, A5411000281, A5411000278, A5411000279, A5411000280, A5411000492, A5411000282, A5411000245, A5411000277, A5411000064, A5411000065, A5411000187, A5411000198, A5411000207, A5411000211, A5411000212, A5411000231, A5411000399, A5411000491, A5411000493. A5411000524, A5411000497, A5411000534, A5411000526, A5411000535, A5411000122, A5411000575, A5411000576, A5411000577, A5411000495, A5411000578, A5410000044, A5411000037, A5411000241, A5411000246, A5411000108, A5411000110, A5411000114, A5411000312, A5411000074, A5411000080, A5411000522, A5411000033, A5411000139, A5411000543, A5411000182, A5411000190, A5411000021, A5411000374, A5411000218, A5411000572, A5411000084, A5411000098, A5411000238, A5411000116, A5411000408, A5411000409, A5410000042, A5410000023, A5411000544, A5411000580, A5411000581, A5411000545, A5411000546, A5411000547, A5411000579, A5411000337, A5411000338, A5411000340, A5411000341, A5411000475, A5411000549, A5411000373, A5411000420, A5410000133, A5410000117, A5410000115, A5410000079, A5410000131, A5410000066, A5410000132, A5410000114, A5410000082, A5410000116, A5411000025, A5411000523, A5411000525, A5411000590, A5411000592, A5411000263, A5411000309, A5411000262, A5411000270, A5411000272, A5411000273, A5411000293, A5411000363, A5411000260, A5411000264, A5411000269, A5411000271, A5411000291, A5411000292, A5411000261, A5411000265, A5411000266, A5411000267, A5411000034, A5411000081, A5411000146, A5411000092, A5411000093, A5411000407, A5411000026, A5411000068, A5411000367, A5411000368, A5411000465, A5411000060, A5411000131, A5411000018, A5411000100, A5411000515, A5411000313, A5411000284, A5411000076, A5411000135, A5411000180, A5411000343, A5411000285, A5411000311, A5411000375, A5411000062, A5411000136, A5411000179, A5411000178, A5411000421, A5411000427, A5411000
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAAgfa Healthcare NX 3.0.8950 Imaging Processing Software26 de julio de 2017
- FDAIMPAX Cardiovascular The IMPAX CV Reporting module consists of a database and graphical user interface (GUI)5 de julio de 2017
- FDADX-D600 - DXD Imaging Package12 de abril de 2017
- FDAProsound F75 or F75 The Hitachi Aloka Medical, Ltd. Prosound F75 is intended for use by trained personnel5 de abril de 2017
- FDAThe Stryker Visum® LED Surgical Lighting system is intended to illuminate the operative site during surgical5 de abril de 2017
- FDARadiomat M+ NIF 14 x 17, Radiomat M+ NIF 35CM x 43CM, Radiomat M+ NIF 10 x 12, Radiomat M+ NIF 25,4CM x 30,5CM24 de diciembre de 2014
Preguntas sobre este retiro
¿AGFA Digital Radiography X-Ray System DX-D100 DX-D100 is indicated for use in providing diagnostic quality images to aid the physician with diagnosis fue retirado del mercado?
Sí. FDA publicó el retiro Z-1698-2017 el 5 de abril de 2017. Peligro reportado: The DX-D100 User Manual already contained information about to move a DX-D100 mobile unit manually by using an Allen wrench, but did not indicate where the Allen wrench should be stored.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter