Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1711-2017

UniCel DxC 600i SYNCHRON Access Clinical System, Part No. A25637, A25638 Product Usage: The Access 2 and UniCel DxI Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids.

Publicado el 12 de abril de 2017 · Inicio del retiro: 7 de marzo de 2017

Resumen

FDA registra el retiro Z-1711-2017 de dispositivos médicos de Clase II, publicado el 12 de abril de 2017 e iniciado el 7 de marzo de 2017. Empresa responsable: Beckman Coulter Inc.; Estado: Finalizado; Cantidad reportada: 851 units total for all products (291 units in US); Lote: Serial No. (Part No. A25637) 800999,801019 Serial No. (Part No. A25638)800995,800996,800997,800998,801000,801001,801002,801003,801004, 801005,801006,801007,801008,801009,801010,801011,801012,801013, 801014,801015,801016,801017,801018,801020,801021,801022,801023, 801024,801025,801026,801027. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Beckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position errors cause the instrument to stop running and cancel tests that are in process.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Beckman Coulter Inc.
Distribuidor
Affected geographies: Algeria, Austria, Australia, Bangladesh, Bulgaria, Brazil, Canada, China, Costa Rica, Cyprus, Colombia, Czech Republic, El Salvador, France, Germany, Ghana, Greece, Guatemala, India, Indonesia, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Lithuania, Malawi, Mexico, Morocco, Myanmar, Nepal, Oman, Panama, Philippines, Pakistan, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Switzerland, Tunisia, Turkey, United Kingdom, United States, Vietnam, Zimbabwe
Lote
Serial No. (Part No. A25637) 800999,801019 Serial No. (Part No. A25638)800995,800996,800997,800998,801000,801001,801002,801003,801004, 801005,801006,801007,801008,801009,801010,801011,801012,801013, 801014,801015,801016,801017,801018,801020,801021,801022,801023, 801024,801025,801026,801027
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿UniCel DxC 600i SYNCHRON Access Clinical System, Part No. A25637, A25638 Product Usage: The Access 2 and UniCel DxI Access Immunoassay Systems are in vitro diagnostic devices used for the quantitative, semi-quantitative or qualitative determination of various analyte concentrations found in human body fluids. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1711-2017 el 12 de abril de 2017. Peligro reportado: Beckman Coulter is recalling the Access 2 Family of Immunoassay Systems because a circuit board that controls motor functions can affect device positioning on the pipettor and incubator belt while the instrument is running. The position errors cause the instrument to stop running and cancel tests that are in process.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter