HLD Systems Model 610 Washer/High Level Disinfector. Maximum weight of the Model 610 when filled with water is approximately 450 pounds. The dimensions of the system are: Width: 26.5" (67 cm); Depth: 33" (83 cm); Height: 46.5" (118 cm) to top of control panel; 64" (162.6 cm)- height clearance with lid open. Product Usage: Usage: This device washes reusable medical devices and disinfects them with high temperature hot water (pasteurization). The devices are washed with a cleaning solution (containing NaClO (sodium hypochlorite), surfactants, and detergents) to aid in the removal of debris and visible stains. Next, the medical devices and rinsed two times with water. The final, disinfection step uses only hot water at 160 F ¿ 2 F for 30 minutes.
Publicado el 18 de junio de 2014 · Inicio del retiro: 24 de abril de 2014
Resumen
FDA registra el retiro Z-1729-2014 de dispositivos médicos de Clase II, publicado el 18 de junio de 2014 e iniciado el 24 de abril de 2014. Empresa responsable: Cenorin, LLC; Estado: Finalizado; Cantidad reportada: 25 HLD systems; Lote: 610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 610902, 611001, 611002, 611004, 611102, 611103, 611201, 611202, 611203, 610902L, 611204. First two digits (61) refer to model number 610 for the HLD system Second two digits - two digit year of manufacture Last two digits sequentially numbered as they receive orders. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Cenorin LLC is voluntarily recalling HLD Systems Model 610 Washer/High Level Disinfector due to a potential risk of the system not injecting the proper amount of cleaning fluid during the wash cycle and not achieving the desired level of debris reduction.
Qué hacer
Voluntary: Firm initiated · E-Mail
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Cenorin, LLC
- Distribuidor
- Distribution USA Nationwide
- Lote
- 610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 610902, 611001, 611002, 611004, 611102, 611103, 611201, 611202, 611203, 610902L, 611204. First two digits (61) refer to model number 610 for the HLD system Second two digits - two digit year of manufacture Last two digits sequentially numbered as they receive orders.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAThermaSure(TM), a CENORIN(TM) product, SERIES 300 MEDICAL DEVICE DRYER: (a) Model 330; (b) Model 340 Forced23 de agosto de 2017
- FDAThermaSure(TM), a CENORIN(TM) product, SERIES 1000 MEDICAL DEVICE DRYER: (a) Model 1009; (b) Model 1010 Forced23 de agosto de 2017
- FDAHLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.16 de julio de 2014
- FDAHorizon Medical Imaging Horizon Medical Imaging is a medical image and information management application that18 de junio de 2014
- FDAMaquet PowerLED 700. Product Usage; To provide illumination to the surgical area or the patient.18 de junio de 2014
- FDAAlere Triage® D-Dimer Test PN 98100, Lot W53884B. The Alere Triage® D-Dimer Test is a fluorescence immunoassay18 de junio de 2014
Preguntas sobre este retiro
¿HLD Systems Model 610 Washer/High Level Disinfector. Maximum weight of the Model 610 when filled with water is approximately 450 pounds. The dimensions of the system are: Width: 26.5" (67 cm); Depth: 33" (83 cm); Height: 46.5" (118 cm) to top of control panel; 64" (162.6 cm)- height clearance with lid open. Product Usage: Usage: This device washes reusable medical devices and disinfects them with high temperature hot water (pasteurization). The devices are washed with a cleaning solution (containing NaClO (sodium hypochlorite), surfactants, and detergents) to aid in the removal of debris and visible stains. Next, the medical devices and rinsed two times with water. The final, disinfection step uses only hot water at 160 F ¿ 2 F for 30 minutes. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1729-2014 el 18 de junio de 2014. Peligro reportado: Cenorin LLC is voluntarily recalling HLD Systems Model 610 Washer/High Level Disinfector due to a potential risk of the system not injecting the proper amount of cleaning fluid during the wash cycle and not achieving the desired level of debris reduction.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · E-Mail