Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1758-2024

PERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55 30-S TLF-DS 55 35-S TLF-DS 55 40-S TLF-DS 55 45-S TLF-DS 65 35-S TLF-DS 65 40-S TLF-DS 65 45-S TLF-DS 65 50-S The PERLA® TL system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (painful degeneration of the disc), spondylolisthesis, trauma, spinal stenosis, deformities (i.e. scoliosis, kyphosis, or lordosis), tumor and failed previous fusion (pseudarthrosis).

Publicado el 15 de mayo de 2024 · Inicio del retiro: 15 de marzo de 2024

Resumen

FDA registra el retiro Z-1758-2024 de dispositivos médicos de Clase II, publicado el 15 de mayo de 2024 e iniciado el 15 de marzo de 2024. Empresa responsable: SPINEART SA; Estado: En curso; Cantidad reportada: 242 systems; Lote: Model Numbers/Lot-Batch Numbers: TLF-DS 45 25-S 7-4407 TLF-DS 45 25-S 7-6207 TLF-DS 45 30-S 7-4244 TLF-DS 45 30-S 7-6226 TLF-DS 45 35-S 7-4242 TLF-DS 45 35-S 7-5470 TLF-DS 45 35-S 7-6684 TLF-DS 45 40-S 7-6232 TLF-DS 45 45-S 7-5474 TLF-DS 45 45-S 7-6874 TLF-DS 55 30-S 7-4409 TLF-DS 55 30-S 7-4695 TLF-DS 55 30-S 7-4965 TLF-DS 55 35-S 7-4413 TLF-DS 55 35-S 7-6455 TLF-DS 55 40-S 7-4415 TLF-DS 55 40-S 7-4417 TLF-DS 55 45-S 7-4411 TLF-DS 55 45-S 7-5055 TLF-DS 55 45-S 7-6087 TLF-DS 55 45-S 7-6459 TLF-DS 65 35-S 7-4405 TLF-DS 65 35-S 7-6230 TLF-DS 65 35-S 7-6461 TLF-DS 65 35-S 7-6882 TLF-DS 65 40-S 7-4248 TLF-DS 65 40-S 7-4250 TLF-DS 65 40-S 7-5480 TLF-DS 65 40-S 7-6228 TLF-DS 65 40-S 7-6554 TLF-DS 65 40-S 7-6740 TLF-DS 65 45-S 7-4582 TLF-DS 65 45-S 7-5484 TLF-DS 65 45-S 7-5808 TLF-DS 65 45-S 7-6668 TLF-DS 65 45-S 7-6742 TLF-DS 65 45-S 7-7299 TLF-DS 65 50-S 7-5488 TLF-DS 65 50-S 7-9264 UDI-DI for associated Model Numbers: TLF-DS 45 25-S 07640185345799 TLF-DS 45 30-S 07640185345805 TLF-DS 45 35-S 07640185345812 TLF-DS 45 40-S 07640185345829 TLF-DS 45 45-S 07640185345836 TLF-DS 55 30-S 07640185345850 TLF-DS 55 35-S 07640185345867 TLF-DS 55 40-S 07640185345874 TLF-DS 55 45-S 07640185345881 TLF-DS 60 35-S 07640305166082. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but position askew

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
SPINEART SA
Distribuidor
U.S. Nationwide distribution in the states of CA, and KY. Not provided due to affected devices were manufactured O.U.S.
Lote
Model Numbers/Lot-Batch Numbers: TLF-DS 45 25-S 7-4407 TLF-DS 45 25-S 7-6207 TLF-DS 45 30-S 7-4244 TLF-DS 45 30-S 7-6226 TLF-DS 45 35-S 7-4242 TLF-DS 45 35-S 7-5470 TLF-DS 45 35-S 7-6684 TLF-DS 45 40-S 7-6232 TLF-DS 45 45-S 7-5474 TLF-DS 45 45-S 7-6874 TLF-DS 55 30-S 7-4409 TLF-DS 55 30-S 7-4695 TLF-DS 55 30-S 7-4965 TLF-DS 55 35-S 7-4413 TLF-DS 55 35-S 7-6455 TLF-DS 55 40-S 7-4415 TLF-DS 55 40-S 7-4417 TLF-DS 55 45-S 7-4411 TLF-DS 55 45-S 7-5055 TLF-DS 55 45-S 7-6087 TLF-DS 55 45-S 7-6459 TLF-DS 65 35-S 7-4405 TLF-DS 65 35-S 7-6230 TLF-DS 65 35-S 7-6461 TLF-DS 65 35-S 7-6882 TLF-DS 65 40-S 7-4248 TLF-DS 65 40-S 7-4250 TLF-DS 65 40-S 7-5480 TLF-DS 65 40-S 7-6228 TLF-DS 65 40-S 7-6554 TLF-DS 65 40-S 7-6740 TLF-DS 65 45-S 7-4582 TLF-DS 65 45-S 7-5484 TLF-DS 65 45-S 7-5808 TLF-DS 65 45-S 7-6668 TLF-DS 65 45-S 7-6742 TLF-DS 65 45-S 7-7299 TLF-DS 65 50-S 7-5488 TLF-DS 65 50-S 7-9264 UDI-DI for associated Model Numbers: TLF-DS 45 25-S 07640185345799 TLF-DS 45 30-S 07640185345805 TLF-DS 45 35-S 07640185345812 TLF-DS 45 40-S 07640185345829 TLF-DS 45 45-S 07640185345836 TLF-DS 55 30-S 07640185345850 TLF-DS 55 35-S 07640185345867 TLF-DS 55 40-S 07640185345874 TLF-DS 55 45-S 07640185345881 TLF-DS 60 35-S 07640305166082
Clasificación de riesgo
Class II
Área (FDA)
device
País
Switzerland

Retiros relacionados

Preguntas sobre este retiro

¿PERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55 30-S TLF-DS 55 35-S TLF-DS 55 40-S TLF-DS 55 45-S TLF-DS 65 35-S TLF-DS 65 40-S TLF-DS 65 45-S TLF-DS 65 50-S The PERLA® TL system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (painful degeneration of the disc), spondylolisthesis, trauma, spinal stenosis, deformities (i.e. scoliosis, kyphosis, or lordosis), tumor and failed previous fusion (pseudarthrosis). fue retirado del mercado?

Sí. FDA publicó el retiro Z-1758-2024 el 15 de mayo de 2024. Peligro reportado: Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but position askew

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter