Welch Allyn ProBP 2400 Digital Blood Pressure Device #2400, REF 901096, Rx ONLY, Made in China --- Distributed by Welch Allyn Inc., 4341 State Street Road, Skaneateles Falls, NY 13153 --- Microlife Corporation 9F, 431, RuiGuang Road, NeiHu, Taipei, 11492, Taiwan, R.O.C. --- Microlife AG Espenstrasse 139 9443 Widnau, Switzerland
Publicado el 19 de abril de 2017 · Inicio del retiro: 27 de febrero de 2017
Resumen
FDA registra el retiro Z-1777-2017 de dispositivos médicos de Clase II, publicado el 19 de abril de 2017 e iniciado el 27 de febrero de 2017. Empresa responsable: Welch Allyn Inc; Estado: Finalizado; Cantidad reportada: 981 units; Lote: Lot Number 0715: Serial Number range (21)07150001 - (21)07150620; Lot Number 1215: Serial Number range (21)12150001 - (21)12150500; --- Note: (21) is NOT part of the serial number. It is the Global Unique Device Identification (GUDI) Application Identifier that precedes the serial number on any GUDI compliant label. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
A defective electrical component in the battery charging circuit in two lots of the ProBP 2400 may potentially cause an over-voltage battery charging condition when the device is connected to the external power supply/battery charger. This potential defect resides with the device, not the battery. The over-voltage condition can result in high NiMH battery temperatures that, in some cases, can reach levels sufficient to cause melting of the plastic battery door and other adjacent plastic and foam device components.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Welch Allyn Inc
- Distribuidor
- Worldwide Distribution - USA (nationwide) Distribution
- Lote
- Lot Number 0715: Serial Number range (21)07150001 - (21)07150620; Lot Number 1215: Serial Number range (21)12150001 - (21)12150500; --- Note: (21) is NOT part of the serial number. It is the Global Unique Device Identification (GUDI) Application Identifier that precedes the serial number on any GUDI compliant label.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACombi Dockable Neurosurgery table for the Siemens MAGNETOM Aera (1.5T) and Skyra (3T) MRI Systems19 de abril de 2017
- FDAMerge Hemo software. The firm name on the labeling is Merge Healthcare, Hartland, WI. Merge Hemo monitors,19 de abril de 2017
- FDACustom Tubing Pack with Bioline Coating UDI code: 00607567208230 Part Number: 709000058 The HLM Tubing Sets19 de abril de 2017
- FDAVivid E95/E90/E80, H45581DC, H45581LB and H45581DA.19 de abril de 2017
- FDABedside monitor CSM-1901(Life Scope G9) with main unit CU-192RA. The Problem only affects CSM-1901(Life Scope19 de abril de 2017
- FDAVivid S60/S70/S60N/S70N, H45041SU, H45041SW, H45581MS, H45581PD.19 de abril de 2017
Preguntas sobre este retiro
¿Welch Allyn ProBP 2400 Digital Blood Pressure Device #2400, REF 901096, Rx ONLY, Made in China --- Distributed by Welch Allyn Inc., 4341 State Street Road, Skaneateles Falls, NY 13153 --- Microlife Corporation 9F, 431, RuiGuang Road, NeiHu, Taipei, 11492, Taiwan, R.O.C. --- Microlife AG Espenstrasse 139 9443 Widnau, Switzerland fue retirado del mercado?
Sí. FDA publicó el retiro Z-1777-2017 el 19 de abril de 2017. Peligro reportado: A defective electrical component in the battery charging circuit in two lots of the ProBP 2400 may potentially cause an over-voltage battery charging condition when the device is connected to the external power supply/battery charger. This potential defect resides with the device, not the battery. The over-voltage condition can result in high NiMH battery temperatures that, in some cases, can reach levels sufficient to cause melting of the plastic battery door and other adjacent plastic and foam device components.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter