Bedside monitor CSM-1901(Life Scope G9) with main unit CU-192RA. The Problem only affects CSM-1901(Life Scope G9) communicating with CNS-6201A (PU-621RA) and CNS-9701A (MU-971RA).
Publicado el 19 de abril de 2017 · Inicio del retiro: 13 de marzo de 2017
Resumen
FDA registra el retiro Z-1768-2017 de dispositivos médicos de Clase II, publicado el 19 de abril de 2017 e iniciado el 13 de marzo de 2017. Empresa responsable: Nihon Kohden America Inc; Estado: Finalizado; Cantidad reportada: 654 units total (230 units in US); Lote: Lot/Unit Codes: G9 Serial Number: 230 (UDI: 01049319211121682100230) Serial Number: 232 (UDI: 01049319211121682100232) Life Scope G9 with version 01-19, 01-20, and 01-21 communicating with CNS-6201A (PU-621RA) with version 01-06, 02-10, or 02-40 software. Life Scope G9 with version 01-19, 01-20, and 01-21 communicating with CNS-9701A (MU-971RA) with version 01-89 or lower. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The Pause function on central monitors will not automatically resume when connected to a Life Scope G9 patient monitor.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Nihon Kohden America Inc
- Distribuidor
- Worldwide - US Nationwide Distribution including the countries of Germany, Spain, Brazil, Paraguay, Bolivia, Chile, Colombia, Costa Rica, Mexico, Peru, Pakistan, Israel, Qatar, Saudi Arabia, South Africa, Tunisia, United Arab Emirates, Korea, Malaysia, Singapore, Thailand
- Lote
- Lot/Unit Codes: G9 Serial Number: 230 (UDI: 01049319211121682100230) Serial Number: 232 (UDI: 01049319211121682100232) Life Scope G9 with version 01-19, 01-20, and 01-21 communicating with CNS-6201A (PU-621RA) with version 01-06, 02-10, or 02-40 software. Life Scope G9 with version 01-19, 01-20, and 01-21 communicating with CNS-9701A (MU-971RA) with version 01-89 or lower.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAAdult Cap-ONE Biteblock REF YG-227T The cap-ONE Biteblock is intended to sample CO2 and administer13 de noviembre de 2024
- FDANihon Kohden Adult Ear Clip SpO2 Sensor, 1.5 meter Model 809030007 Reusable SpO2 Ear Clip, each6 de noviembre de 2024
- FDANihon Kohden Adult/Pediatric Forehead Disposable SpO2 Sensor, 0.9m Model 809030006 Disp. SpO2 Forehead Probe,6 de noviembre de 2024
- FDABSM-3000 Series Bedside Monitor REF BSM-3572A The Life Scope BSM-3000 Series Bedside Monitoring System is12 de junio de 2024
- FDACADD® Administration Set, 21-7339-24, Rx Only, Sterile EO.19 de abril de 2017
- FDACustom Tubing Pack with Bioline Coating UDI code: 00607567208230 Part Number: 709000058 The HLM Tubing Sets19 de abril de 2017
Preguntas sobre este retiro
¿Bedside monitor CSM-1901(Life Scope G9) with main unit CU-192RA. The Problem only affects CSM-1901(Life Scope G9) communicating with CNS-6201A (PU-621RA) and CNS-9701A (MU-971RA). fue retirado del mercado?
Sí. FDA publicó el retiro Z-1768-2017 el 19 de abril de 2017. Peligro reportado: The Pause function on central monitors will not automatically resume when connected to a Life Scope G9 patient monitor.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter