Convatec Flexi Seal Control Fecal Management System Rx Only REF 411107 Product Usage:For use to manage fecal incontinence through the collection of liquid to semi-liquid stool and to provide access to administer medications as prescribed by a physician.
Publicado el 25 de junio de 2014 · Inicio del retiro: 29 de abril de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Convatec Inc. is recalling Flexi-seal FMS Control Kit since it does not have a 510K. In addition, an internal assessment of product performance, including a review of customer feedback, has confirmed that Flexi-Seal Control is not meeting expectations or those of customers. Specifically, the Auto-Valve feature that is unique to the Flexi-Seal Control has not been consistently performing relative
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Convatec Inc.
- Distribuidor
- Worldwide Distribution - USA Nationwide
- Lote
- 13-FM-0301 13-FM-0302 13FM0303 13-FM-0303 13FM0309 13VM509948 13VM509949 13VM510046 13VM512473 13VM513636 13VM513642 13VM516733 13VM520039 13VM521731 13VM521733 13VM524363 13VM524364 13VM527149 13VM527150 13VM528992 13VM528994 13VM531214 14FM0301 14VM531935 14VM532098
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States