FDAFDA · Alimentos y medicamentos·N.º oficial Z-1788-2013

GE Healthcare Varicam/VG and VG Hawkeye devices Product Usage: The intended use of the Quasar system is to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition typed and optional imaging features designed to enhance image quality in Oncology, Cardiology, Neurology and other clinical diagnostic imaging applications. The scanning modes include planar (Static, Multi-gated, Dynamic, Whole body scanning) and tomographic (SPECT, Gated SPECT, Whole body SPECT, Camera based PET). Acquisition types include single and multi-isotope/multi-peak frame/list mode single-photon and positron imaging. Optional imaging-enhancement features include assortment of collimators, gating by physiological signals, real-time automatic body contouring, and CT-based attenuation correction and functional anatomic mapping.

Publicado el 31 de julio de 2013 · Inicio del retiro: 13 de junio de 2013

Resumen

FDA registra el retiro Z-1788-2013 de dispositivos médicos de Clase I, publicado el 31 de julio de 2013 e iniciado el 13 de junio de 2013. Empresa responsable: GE Healthcare, LLC; Estado: Finalizado; Cantidad reportada: 571; Lote: System ID: 2998, 6970, 12865, 21414, 38070, 53263, 55461, 56948, 58230, 58378, 58379, 58483, 58510, 58582, 58622, 58630, 58711, 58785, 58797, 58808, 58810, 58871, 58883, 58919, 58958, 58993, 59012, 61061, 63422, 66515, 75074, 79722, 79966, 84636, 92635, 99691, 0850260033, 0850260034, 0850260035, 0850260038, 0850260040, 0850260041, 0850260042, 0850260046, 0850260048, 0850260054, 0850260057, 0850260062, 0853940029, 0853940068, 0853940069, 0853940077, 0853940078, 0853940080, 0853940082, 0856260014, 0856260019, 0910265011, 0910453006, 082426010009, 082426010010, 082426010011, 082426020002, 082426030009, 082426030010, 082426030015, 082426030016, 082426030018, 082426040016, 082426040020, 082426040021, 082426040024, 082426040025, 082426040026, 082426040029, 082426050001, 082426050002, 082426050003, 082426070014, 082426070015, 082426070016, 082426080006, 082426080007, 082426080009, 082426090013, 082426100017, 082426100019, 082426100020, 082426100023, 082426110002, 082426120003, 082426120004, 082426130012, 082426130013, 082426130014, 082426130015, 082426130016, 082426130019, 082426140005, 082426140007, 082426140008, 082426140009, 082426140011, 082426140012, 082426150003, 082426150006, 082426160012, 082426160014, 082426160016, 082426160017, 082426160019, 082426160020, 082426160022, 082426160023, 082426180003, 082426190008, 082426190010, 082426200002, 082426210002, 082426210003, 082426210010, 082426220006, 082426230008, 082426230010, 082426250002, 082426280001, 082426290004, 082426310005, 082426310007, 082426310008, 08302603017002, 08302603135243, 08302606123002, 08302621006334, 083026031062021, 083026242100654, 083026590040004, 083026590040010, 083026598000594, 083026826753913, 0382E00013, 00023NUC06, 0009D, 00132NUC02, 00216NUC09, 00243NUC11, 00243NUC15, 00243NUC17, 00269NUC07, 00323NUC02, 00357NUC02, 00555NUC01, 00584NUC01, 00584NUC04, 0112B, 0277A, 030010NU02, 030028NU01, 030036NU06, 030082NU10, 030151NU05, 030220NU02, 030355NU02, 030388NU01, 0306A, 0343A, 0360A, 06007NUC16, 080041NU01, 082426040029P, 082426050001P, 082426140005P, 0850EL0011, 0850EL0012, 10433NUC01, 110012NU02, 110965NU05, 130083NU07, 13228MILLEN, 16003NUC03, 160070NU01, 190226NU17, 190282NU04, 203694MIL, 203852VG, 203932VG, 204837VG, 205823NMVG, 205933VH, 206288VG, 206598VG, 207795VH, 208529VG, 209334LNVG1, 209526MILVG, 212562VG, 215893VG, 220166NU01, 223410MILVG, 225763MVG, 228818MILLVG, 248828H, 248828VG, 248937VG, 256235OPNM1, 262785VG, 2894872VG2, 2894872VG3, 301896VG, 304256VG, 304487VG, 309662SJVG, 310231VGH1, 310289VG1, 312864VG, 313299VG, 334794MILVG, 361575VG, 402449SJVG, 404605PVGH, 404712VG, 406782SNM, 407277FVG2, 407896FVG1, 407933KVG, 408947VG, 409291HVG, 410605NM1, 410955VG1, 410955VH, 412359VG, 413781VG1, 413781VG2, 414219VG, 414328VG, 414384VG, 414447VG, 414544HEYE, 414544VG, 414777MIL, 414777VG2, 414961VG, 416284VG, 416284VG2, 416290VG, 416444VG, 416469DVH, 416534VG1, 416534VG2, 416586VG, 416656VG, 416747VARICAM, 416747VG, 416756VG, 418404NU01, 418862VARICAM, 419383VG, 440329VG, 443643VG, 450359VG1, 450359VG2, 480981NM1, 501686VG, 508997VG, 509575PYVGH, 510486VG1, 512753VG, 514495VG, 515263LVG, 516256VG, 518271SVG, 519421VG, 520757NM1, 530529VG, 540186NU01, 540596N3, 540953VG, 541276AVG, 559688NM1, 570791HAWK, 580355VGACQ, 589907MVG, 600062NM01, 601426VG, 602406VG, 602870NM1, 605333HAWKEYE, 608263VG, 608364VG, 609267VG, 610954VG, 613548VG, 613548VG2, 615322NM2, 615791VG, 630801VG, 70050NUC06, 705324VG, 705327VARICAM, 705743VG, 706845VG, 707554VG, 708422MIL, 708783VG, 714774VG, 715361VG, 716694VG, 716859VG, 717657VG, 718759VG, 718780GENIEVG, 718780VG, 73240-1, 781278VG, 781744VG2, 781744VH, 786090NU03, 787782VARICAM-1, 804281VG1, 807343BMILLVG, 807343CMILLVG, 809758VARICAM, 812537VG, 813725VG, 813788VG, 815842SJVG, 819563CVG, 819822VG, 828212VG, 828213VG, 843792NUCH, 845437VG, 847695MILMVG, 847842VG, 860343VGOPC, 860344VG, 863494VG, 864573MBVG1, 904202VG, 905472VG, 905527HAWKEYE, 905683VG, 905684VG, 905684VG2, 905845VG, 905848VG, 905855VG, 905883VG,. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
GE Healthcare, LLC
Distribuidor
Worldwide Distribution - US Nationwide including Guam, Puerto Rico and the countries of : Afghanistan, Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Korea (Republic of), Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Malta, Martinique, Mexico, Morocco, Namibia, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, Saudi Arabia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, Uruguay, Venezuela, Vietnam.
Lote
System ID: 2998, 6970, 12865, 21414, 38070, 53263, 55461, 56948, 58230, 58378, 58379, 58483, 58510, 58582, 58622, 58630, 58711, 58785, 58797, 58808, 58810, 58871, 58883, 58919, 58958, 58993, 59012, 61061, 63422, 66515, 75074, 79722, 79966, 84636, 92635, 99691, 0850260033, 0850260034, 0850260035, 0850260038, 0850260040, 0850260041, 0850260042, 0850260046, 0850260048, 0850260054, 0850260057, 0850260062, 0853940029, 0853940068, 0853940069, 0853940077, 0853940078, 0853940080, 0853940082, 0856260014, 0856260019, 0910265011, 0910453006, 082426010009, 082426010010, 082426010011, 082426020002, 082426030009, 082426030010, 082426030015, 082426030016, 082426030018, 082426040016, 082426040020, 082426040021, 082426040024, 082426040025, 082426040026, 082426040029, 082426050001, 082426050002, 082426050003, 082426070014, 082426070015, 082426070016, 082426080006, 082426080007, 082426080009, 082426090013, 082426100017, 082426100019, 082426100020, 082426100023, 082426110002, 082426120003, 082426120004, 082426130012, 082426130013, 082426130014, 082426130015, 082426130016, 082426130019, 082426140005, 082426140007, 082426140008, 082426140009, 082426140011, 082426140012, 082426150003, 082426150006, 082426160012, 082426160014, 082426160016, 082426160017, 082426160019, 082426160020, 082426160022, 082426160023, 082426180003, 082426190008, 082426190010, 082426200002, 082426210002, 082426210003, 082426210010, 082426220006, 082426230008, 082426230010, 082426250002, 082426280001, 082426290004, 082426310005, 082426310007, 082426310008, 08302603017002, 08302603135243, 08302606123002, 08302621006334, 083026031062021, 083026242100654, 083026590040004, 083026590040010, 083026598000594, 083026826753913, 0382E00013, 00023NUC06, 0009D, 00132NUC02, 00216NUC09, 00243NUC11, 00243NUC15, 00243NUC17, 00269NUC07, 00323NUC02, 00357NUC02, 00555NUC01, 00584NUC01, 00584NUC04, 0112B, 0277A, 030010NU02, 030028NU01, 030036NU06, 030082NU10, 030151NU05, 030220NU02, 030355NU02, 030388NU01, 0306A, 0343A, 0360A, 06007NUC16, 080041NU01, 082426040029P, 082426050001P, 082426140005P, 0850EL0011, 0850EL0012, 10433NUC01, 110012NU02, 110965NU05, 130083NU07, 13228MILLEN, 16003NUC03, 160070NU01, 190226NU17, 190282NU04, 203694MIL, 203852VG, 203932VG, 204837VG, 205823NMVG, 205933VH, 206288VG, 206598VG, 207795VH, 208529VG, 209334LNVG1, 209526MILVG, 212562VG, 215893VG, 220166NU01, 223410MILVG, 225763MVG, 228818MILLVG, 248828H, 248828VG, 248937VG, 256235OPNM1, 262785VG, 2894872VG2, 2894872VG3, 301896VG, 304256VG, 304487VG, 309662SJVG, 310231VGH1, 310289VG1, 312864VG, 313299VG, 334794MILVG, 361575VG, 402449SJVG, 404605PVGH, 404712VG, 406782SNM, 407277FVG2, 407896FVG1, 407933KVG, 408947VG, 409291HVG, 410605NM1, 410955VG1, 410955VH, 412359VG, 413781VG1, 413781VG2, 414219VG, 414328VG, 414384VG, 414447VG, 414544HEYE, 414544VG, 414777MIL, 414777VG2, 414961VG, 416284VG, 416284VG2, 416290VG, 416444VG, 416469DVH, 416534VG1, 416534VG2, 416586VG, 416656VG, 416747VARICAM, 416747VG, 416756VG, 418404NU01, 418862VARICAM, 419383VG, 440329VG, 443643VG, 450359VG1, 450359VG2, 480981NM1, 501686VG, 508997VG, 509575PYVGH, 510486VG1, 512753VG, 514495VG, 515263LVG, 516256VG, 518271SVG, 519421VG, 520757NM1, 530529VG, 540186NU01, 540596N3, 540953VG, 541276AVG, 559688NM1, 570791HAWK, 580355VGACQ, 589907MVG, 600062NM01, 601426VG, 602406VG, 602870NM1, 605333HAWKEYE, 608263VG, 608364VG, 609267VG, 610954VG, 613548VG, 613548VG2, 615322NM2, 615791VG, 630801VG, 70050NUC06, 705324VG, 705327VARICAM, 705743VG, 706845VG, 707554VG, 708422MIL, 708783VG, 714774VG, 715361VG, 716694VG, 716859VG, 717657VG, 718759VG, 718780GENIEVG, 718780VG, 73240-1, 781278VG, 781744VG2, 781744VH, 786090NU03, 787782VARICAM-1, 804281VG1, 807343BMILLVG, 807343CMILLVG, 809758VARICAM, 812537VG, 813725VG, 813788VG, 815842SJVG, 819563CVG, 819822VG, 828212VG, 828213VG, 843792NUCH, 845437VG, 847695MILMVG, 847842VG, 860343VGOPC, 860344VG, 863494VG, 864573MBVG1, 904202VG, 905472VG, 905527HAWKEYE, 905683VG, 905684VG, 905684VG2, 905845VG, 905848VG, 905855VG, 905883VG,
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿GE Healthcare Varicam/VG and VG Hawkeye devices Product Usage: The intended use of the Quasar system is to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition typed and optional imaging features designed to enhance image quality in Oncology, Cardiology, Neurology and other clinical diagnostic imaging applications. The scanning modes include planar (Static, Multi-gated, Dynamic, Whole body scanning) and tomographic (SPECT, Gated SPECT, Whole body SPECT, Camera based PET). Acquisition types include single and multi-isotope/multi-peak frame/list mode single-photon and positron imaging. Optional imaging-enhancement features include assortment of collimators, gating by physiological signals, real-time automatic body contouring, and CT-based attenuation correction and functional anatomic mapping. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1788-2013 el 31 de julio de 2013. Peligro reportado: GE became aware of an incident at a VA Medical Center facility in the US in which a patient died due to injuries sustained while being scanned on an Infinia Hawkeye 4 Nuclear Medicine System.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter