Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1834-2013

AU5800 Clinical Chemistry Analyzer with Automation Ready Connection (All Software Versions), Model Numbers AU5811, AU5821, AU5831 and AU5841. This system is for in vitro diagnostic use only.

Publicado el 7 de agosto de 2013 · Inicio del retiro: 3 de junio de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The recall was initiated because Beckman Coulter has identified a software error on AU5800 analyzers with the Automation Ready (AR) Sampler Kit. This software error results in the failure of the concentrated wash solution pump to stop, creating an overflow or leak on each analyzer unit.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Beckman Coulter Inc.
Distribuidor
Worldwide Distribution, including Nationwide (US) and the countries of China, Italy, Japan, Canada, Turkey, United Kingdom, Singapore, Israel, and the Netherlands.
Lote
Model Numbers: AU5811-01, AU5821-01, AU5831-01, AU5841-01, AU5811-02, AU5821-02, AU5831-02, AU5841-02, AU5811-03, AU5821-03, AU5831-03, AU5841-03, AU5811-04, AU5821-04, AU5831-04, AU5841-04, AU5811-06, AU5821-06, AU5831-06, AU5841-06 Part Numbers: A94906, A94911, A94916, A91921 A94907, A94912, A94917, A94922 A94908, A94913, A94918, A94923 A94909, A94914, A91949, A94924 A94910, A94915, A94920, A94925 Serial Numbers: All
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta