The Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationary grid. It contains signaling electronics to indicate the presence of the grid and detector as well as detector orientation. It may be mounted to a wall stand or incorporated in a radiographic table. It is intended to be used in general radiographic applications as a component of a radiographic system.
Publicado el 9 de julio de 2014 · Inicio del retiro: 15 de marzo de 2014
Resumen
FDA registra el retiro Z-1836-2014 de dispositivos médicos de Clase II, publicado el 9 de julio de 2014 e iniciado el 15 de marzo de 2014. Empresa responsable: Claymount Assembli; Estado: Finalizado; Cantidad reportada: 12 devices; Lote: *** US Devices Only *** 1) Part ID: 17360; Serial Numbers: WO186722/E-1 , WO196470/E-1 , WO198534/E-1. 2) Part ID: 17361; Serial Numbers: WO186723/E-1. 3) Part ID: 17362; Serial Numbers: WO186724/E-1 , WO196360/E-1 , WO198545/E-2. 4) Part ID: 18476; Serial Numbers: WO196188/E-1 , WO196188/E-2 , WO196189/E-2. 5) Part ID: 18477; Serial Numbers: WO198547/E-1. 6) Part ID: 18479; Serial Numbers: WO198552/E-1. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Claymount Assemblies BV is recalling the Harmony2 model of Harmony Bucky because 1) the devices are incorrectly labeled in that the label does not list the manufacturer address, the date of manufacture, or a statement of compliance with 21 CFR Subchapter J, and 2) the device's insulation values do not comply with IEC 60601-1 in all cases.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Claymount Assembli
- Distribuidor
- Worldwide Distribution - USA including NC, OH, CO, CA, NJ, IL, OH, PA, FL and Internationally to Canada, China, France, Italy, Netherlands, South Africa, Switzerland, Turkey, and United Kingdom.
- Lote
- *** US Devices Only *** 1) Part ID: 17360; Serial Numbers: WO186722/E-1 , WO196470/E-1 , WO198534/E-1. 2) Part ID: 17361; Serial Numbers: WO186723/E-1. 3) Part ID: 17362; Serial Numbers: WO186724/E-1 , WO196360/E-1 , WO198545/E-2. 4) Part ID: 18476; Serial Numbers: WO196188/E-1 , WO196188/E-2 , WO196189/E-2. 5) Part ID: 18477; Serial Numbers: WO198547/E-1. 6) Part ID: 18479; Serial Numbers: WO198552/E-1.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Netherlands
Retiros relacionados
- FDAThe Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationary9 de julio de 2014
- FDAMedtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, Catalog No. 46913. Intended9 de julio de 2014
- FDAYangban Rice Porridge with Abalone, Net Wt. 10.16 oz (288 g), UPC 8 83298 61617 3, 24 cans/case9 de julio de 2014
- FDAMedtronic Duet External Drainage and Monitoring System, SmartSite Injection Sites, Catalog No. 46914. Intended9 de julio de 2014
- FDAOcean's Catch Jonah Crab Minced Meat Crabmeat , Net WT. 10 lb. plastic bag9 de julio de 2014
- FDAMedtronic Duet External Drainage and Monitoring System, Interlink Injection Sites, Ventricular Catheter,9 de julio de 2014
Preguntas sobre este retiro
¿The Claymount Harmony is a grid cabinet designed to accept a digital x-ray imaging receptor and stationary grid. It contains signaling electronics to indicate the presence of the grid and detector as well as detector orientation. It may be mounted to a wall stand or incorporated in a radiographic table. It is intended to be used in general radiographic applications as a component of a radiographic system. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1836-2014 el 9 de julio de 2014. Peligro reportado: Claymount Assemblies BV is recalling the Harmony2 model of Harmony Bucky because 1) the devices are incorrectly labeled in that the label does not list the manufacturer address, the date of manufacture, or a statement of compliance with 21 CFR Subchapter J, and 2) the device's insulation values do not comply with IEC 60601-1 in all cases.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter