FDAFDA · Alimentos y medicamentos·N.º oficial Z-1842-2024
AchievaXR -Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device Model Number (REF): 781253
Publicado el 29 de mayo de 2024 · Inicio del retiro: 3 de mayo de 2024
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Philips North America Llc
- Distribuidor
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
- Lote
- UDI-DI: N/A Serial Numbers: 35062 35028 35068 35081 35055 35069 35084 35024 35025 35070 35007 35013 35064 35071 35083 35022 35063 35074 35049 35035 35067 35004 35006 35011 35015 35016 35021 35050 35057 35061 35075 35076 35082 35085
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States