MR system 1.5T Marlin- Magnetic Resonance (MR) systems are Medical Electrical Systems indicated for use as a diagnostic device. Model Number (REF): 781483

Publicado el 29 de mayo de 2024 · Inicio del retiro: 3 de mayo de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The g-MDU (global Mains Distribution Unit) L3 terminal connection may become loose creating a hotspot that may cause smoke/fire to alarm in the hospital's technical room, resulting in Smoke and/or fire

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Philips North America Llc
Distribuidor
Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Japan, Lebanon, Libya, Lithuania, Martinique, Netherlands, Norway, Poland, Romania, Slovakia, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Lote
UDI-DI: N/A Serial Numbers: 79010 79008 79006 79009 79011 79003 79004 79002 79012 79005
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States