Stryker Rose Bed Model (FL14E) Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver, to adjust the height and/or surface contour of the bed.The devices listed above also include movable and latchable side rails which can be lowered or raised to help prevent patient fall and subsequent injury. The Beds include an optional Scale (Models: FL20E, FL25E, FL28C, FL28EX) as well as an optional Bed Exit (1 Zone) or Chaperone
Publicado el 4 de julio de 2012 · Inicio del retiro: 17 de mayo de 2012
Resumen
FDA registra el retiro Z-1861-2012 de dispositivos médicos de Clase II, publicado el 4 de julio de 2012 e iniciado el 17 de mayo de 2012. Empresa responsable: Stryker Medical Division of Stryker Corporation; Estado: Finalizado; Cantidad reportada: 999; Lote: Model FL14E: Serial Number QB22371, QB22370, QB22369, QB22368, QB22367, QB21200, QB18330, QB17733, B22524, B22523, B22522, B22521, B22506, B22505, B22504, B22503, B22502, B22501, B22500, B22446, B22445, B22444, B22443, B22442, B22441, B22440, B22439, B22438, B22437, B22436, B22435, B22427, B22426, B22425, B22424, B22423, B22422, B22421, B22420, B22419, B22373, B22372, B22308, B22307, B22306, B22305, B22304, B22303, B22302, B22301, B22300, B22299, B22298, B22297, B22296, B22295, B22294, B22293, B22292, B22291, B22290, B22289, B22247, B22246, B22245, B22244, B22243, B22242, B22241, B22240, B22239, B22238, B22237, B22236, B22235, B22234, B22233, B22232, B22231, B22230, B22229, B22228, B22227, B22226, B22225, B22224, B22223, B22222, B22221, B22220, B22219, B22218, B22217, B22216, B22215, B22214, B22213, B22212, B22211, B22210, B22209, B22208, B22207, B22206, B22205, B22204, B22203, B22202, B22201, B22200, B22199, B22198, B22197, B22196, B22195, B22194, B22193, B22192, B22191, B22190, B22189, B22188, B22187, B22186, B22185, B22184, B22172, B22171, B22170, B22169, B22168, B22167, B22166, B22165, B22164, B22163, B22162, B22161, B22160, B22159, B22158, B22157, B22156, B22120, B22119, B22118, B22117, B22116, B22115, B22114, B22113, B22112, B22111, B22110, B22109, B22108, B22107, B22106, B22105, B22104, B22103, B22102, B22101, B22100, B22099, B22098, B22097, B22096, B22095, B22094, B22093, B22092, B22091, B22090, B22089, B22088, B22087, B22086, B22085, B22084, B22083, B22082, B22081, B22080, B22079, B22078, B22077, B22076, B22075, B22074, B22073, B22072, B22071, B22070, B22069, B22068, B22067, B22066, B22065, B22064, B22063, B22062, B22061, B22060, B22059, B22058, B22057, B22056, B22055, B22054, B22053, B22052, B22051, B22050, B22049, B22048, B22047, B22046, B22045, B22044, B22043, B22042, B22041, B22040, B22039, B22038, B22037, B22036, B22035, B22034, B22033, B22032, B22031, B22030, B22029, B22028, B22027, B22026, B22025, B22024, B22023, B22022, B22021, B22020, B22019, B22018, B22017, B22016, B22015, B22014, B22013, B22012, B21975, B21974, B21973, B21972, B21971, B21970, B21969, B21968, B21967, B21966, B21965, B21964, B21963, B21962, B21961, B21960, B21959, B21958, B21898, B21897, B21896, B21895, B21894, B21893, B21892, B21891, B21890, B21810, B21809, B21808, B21807, B21806, B21805, B21804, B21803, B21787, B21786, B21785, B21784, B21783, B21782, B21781, B21780, B21779, B21744, B21743, B21742, B21741, B21740, B21739, B21738, B21737, B21736, B21735, B21734, B21733, B21732, B21731, B21730, B21729, B21728, B21727, B21726, B21725, B21724, B21723, B21722, B21721, B21720, B21719, B21718, B21717, B21716, B21715, B21714, B21713, B21712, B21711, B21710, B21709, B21708, B21707, B21706, B21705, B21704, B21703, B21702, B21701, B21700, B21699, B21698, B21697, B21696, B21695, B21694, B21634, B21633, B21632, B21631, B21630, B21629, B21628, B21627, B21626, B21625, B21624, B21623, B21622, B21621, B21620, B21619, B21618, B21595, B21594, B21593, B21592, B21591, B21590, B21589, B21588, B21587, B21586, B21585, B21584, B21583, B21582, B21581, B21580, B21579, B21578, B21577, B21576, B21575, B21574, B21549, B21514, B21513, B21512, B21501, B21500, B21499, B21498, B21497, B21496, B21495, B21494, B21493, B21473, B21472, B21471, B21470, B21469, B21468, B21467, B21466, B21465, B21464, B21463, B21385, B21384, B21383, B21382, B21381, B21380, B21379, B21378, B21348, B21347, B21346, B21345, B21344, B21343, B21342, B21341, B21340, B21339, B21338, B21337, B21336, B21335, B21334, B21316, B21315, B21314, B21313, B21312, B21302, B21301, B21300, B21299, B21298, B21297, B21296, B21295, B21294, B21293, B21292, B21291, B21290, B21207, B21206, B21205, B21204, B21203, B21202, B21201, B21199, B21198, B21197, B21196, B21195, B21189, B21188, B21187, B21186, B21185, B21184, B21183, B21182, B21181, B21180, B21143, B21142, B21141, B21140, B21139, B21138, B21137, B21136, B21135, B21134, B21133, B21132, B21131, B21075, B21074, B21073, B21072, B21071, B21070, B20969, B20968, B20967, B2096. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Stryker Medical Division of Stryker Corporation
- Distribuidor
- Worldwide distribution:USA (nationwide) and countries including: Argentina, Bertec, Brasil, Brazil, Canada, Chile, China, India, Latin America, Mexico,Osteonics SA, Polska, Portugal, Puerto Rico, and Singapore.
- Lote
- Model FL14E: Serial Number QB22371, QB22370, QB22369, QB22368, QB22367, QB21200, QB18330, QB17733, B22524, B22523, B22522, B22521, B22506, B22505, B22504, B22503, B22502, B22501, B22500, B22446, B22445, B22444, B22443, B22442, B22441, B22440, B22439, B22438, B22437, B22436, B22435, B22427, B22426, B22425, B22424, B22423, B22422, B22421, B22420, B22419, B22373, B22372, B22308, B22307, B22306, B22305, B22304, B22303, B22302, B22301, B22300, B22299, B22298, B22297, B22296, B22295, B22294, B22293, B22292, B22291, B22290, B22289, B22247, B22246, B22245, B22244, B22243, B22242, B22241, B22240, B22239, B22238, B22237, B22236, B22235, B22234, B22233, B22232, B22231, B22230, B22229, B22228, B22227, B22226, B22225, B22224, B22223, B22222, B22221, B22220, B22219, B22218, B22217, B22216, B22215, B22214, B22213, B22212, B22211, B22210, B22209, B22208, B22207, B22206, B22205, B22204, B22203, B22202, B22201, B22200, B22199, B22198, B22197, B22196, B22195, B22194, B22193, B22192, B22191, B22190, B22189, B22188, B22187, B22186, B22185, B22184, B22172, B22171, B22170, B22169, B22168, B22167, B22166, B22165, B22164, B22163, B22162, B22161, B22160, B22159, B22158, B22157, B22156, B22120, B22119, B22118, B22117, B22116, B22115, B22114, B22113, B22112, B22111, B22110, B22109, B22108, B22107, B22106, B22105, B22104, B22103, B22102, B22101, B22100, B22099, B22098, B22097, B22096, B22095, B22094, B22093, B22092, B22091, B22090, B22089, B22088, B22087, B22086, B22085, B22084, B22083, B22082, B22081, B22080, B22079, B22078, B22077, B22076, B22075, B22074, B22073, B22072, B22071, B22070, B22069, B22068, B22067, B22066, B22065, B22064, B22063, B22062, B22061, B22060, B22059, B22058, B22057, B22056, B22055, B22054, B22053, B22052, B22051, B22050, B22049, B22048, B22047, B22046, B22045, B22044, B22043, B22042, B22041, B22040, B22039, B22038, B22037, B22036, B22035, B22034, B22033, B22032, B22031, B22030, B22029, B22028, B22027, B22026, B22025, B22024, B22023, B22022, B22021, B22020, B22019, B22018, B22017, B22016, B22015, B22014, B22013, B22012, B21975, B21974, B21973, B21972, B21971, B21970, B21969, B21968, B21967, B21966, B21965, B21964, B21963, B21962, B21961, B21960, B21959, B21958, B21898, B21897, B21896, B21895, B21894, B21893, B21892, B21891, B21890, B21810, B21809, B21808, B21807, B21806, B21805, B21804, B21803, B21787, B21786, B21785, B21784, B21783, B21782, B21781, B21780, B21779, B21744, B21743, B21742, B21741, B21740, B21739, B21738, B21737, B21736, B21735, B21734, B21733, B21732, B21731, B21730, B21729, B21728, B21727, B21726, B21725, B21724, B21723, B21722, B21721, B21720, B21719, B21718, B21717, B21716, B21715, B21714, B21713, B21712, B21711, B21710, B21709, B21708, B21707, B21706, B21705, B21704, B21703, B21702, B21701, B21700, B21699, B21698, B21697, B21696, B21695, B21694, B21634, B21633, B21632, B21631, B21630, B21629, B21628, B21627, B21626, B21625, B21624, B21623, B21622, B21621, B21620, B21619, B21618, B21595, B21594, B21593, B21592, B21591, B21590, B21589, B21588, B21587, B21586, B21585, B21584, B21583, B21582, B21581, B21580, B21579, B21578, B21577, B21576, B21575, B21574, B21549, B21514, B21513, B21512, B21501, B21500, B21499, B21498, B21497, B21496, B21495, B21494, B21493, B21473, B21472, B21471, B21470, B21469, B21468, B21467, B21466, B21465, B21464, B21463, B21385, B21384, B21383, B21382, B21381, B21380, B21379, B21378, B21348, B21347, B21346, B21345, B21344, B21343, B21342, B21341, B21340, B21339, B21338, B21337, B21336, B21335, B21334, B21316, B21315, B21314, B21313, B21312, B21302, B21301, B21300, B21299, B21298, B21297, B21296, B21295, B21294, B21293, B21292, B21291, B21290, B21207, B21206, B21205, B21204, B21203, B21202, B21201, B21199, B21198, B21197, B21196, B21195, B21189, B21188, B21187, B21186, B21185, B21184, B21183, B21182, B21181, B21180, B21143, B21142, B21141, B21140, B21139, B21138, B21137, B21136, B21135, B21134, B21133, B21132, B21131, B21075, B21074, B21073, B21072, B21071, B21070, B20969, B20968, B20967, B2096
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAProCuity bed series, model number 3009, item number: 3009PX-LE-400.13 de diciembre de 2023
- FDAProform- support surface is used to assist in the prevention and treatment of all categories/stages of18 de agosto de 2021
- FDAPrime Fifth Wheel Stretcher-stretcher may be used for minor procedures and short-term stay (treatment and18 de agosto de 2021
- FDAThe Spirit Plus Bed, Model 5600 Spirit Plus /with high rail siderail option21 de noviembre de 2018
- FDAGo Bed II+ Model FL28C Beds are intended for medical purposes in both general patient care as well as critical4 de julio de 2012
- FDAStryker MA204 Bed Model (FL25E) Beds are intended for medical purposes in both general patient care as well as4 de julio de 2012
Preguntas sobre este retiro
¿Stryker Rose Bed Model (FL14E) Beds are intended for medical purposes in both general patient care as well as critical care environments. Each uses built-in electric motors and controls which can be operated by the patient, or caregiver, to adjust the height and/or surface contour of the bed.The devices listed above also include movable and latchable side rails which can be lowered or raised to help prevent patient fall and subsequent injury. The Beds include an optional Scale (Models: FL20E, FL25E, FL28C, FL28EX) as well as an optional Bed Exit (1 Zone) or Chaperone fue retirado del mercado?
Sí. FDA publicó el retiro Z-1861-2012 el 4 de julio de 2012. Peligro reportado: An adverse trend of scale and bed exit service reports/complaints were detected for the GoBed+, GoBed II, Rose, and MA204 model beds. Investigation found an abnormally high failure rate of loadcells, an integral component in scale and bed exit function. The scale faults are manifested as a fluctuating scale reading, a scale error causing the scale system to be disabled until serviced, or an inac
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter