Incisive CT - Computed Tomography X-Ray System intended to produce cross-sectional images of the body. Model Number: 728143

Publicado el 29 de mayo de 2024 · Inicio del retiro: 15 de abril de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Metal mounting box on the rotating scanner on rotor (heat change box) located within the Incisive CT system, may become compromised, resulting in potential contact with other gantry components located within the Incisive CT system during rotation. Resulting in a fragment of a damaged component expelled at a low velocity

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Philips North America Llc
Distribuidor
Worldwide - US Nationwide distribution.
Lote
UDI-DI: (01) 00884838085015 (21) + Serial number Serial Numbers- Applied to FCO72800819: 500046, 500064, 500074, 500102, 500110, 500111, 500145, 500168, 500178, 500211 500216, 500217, 500222, 500225, 500230, 500240, 500242. Serial Number-Applied to FCO72800820 : 500615
Clasificación de riesgo
Class II