FDAFDA · Alimentos y medicamentos·N.º oficial Z-1897-2012

Ethicon Surgiflo Hemostatic Matrix Kit with Thrombin Plus Flex Tip Distributed by Ethicon Ethicon Inc., Somerville, NJ 08876 USA Manufactured by Ferrosan Medical Devices Ferrosan Medical Devices A/S Sydmarken 5, DK-2860 Soeberg, Denmark Mixed with sterile saline or thrombin solution is indicated for surgical procedures (except opthalmic) for hemostasis when control of capillary, venous and arteriolar bleeding by pressure, ligature, and other conventional procedures is ineffective or impractical.

Publicado el 4 de julio de 2012 · Inicio del retiro: 3 de abril de 2012

Resumen

FDA registra el retiro Z-1897-2012 de dispositivos médicos de Clase II, publicado el 4 de julio de 2012 e iniciado el 3 de abril de 2012. Empresa responsable: Ethicon, Inc.; Estado: Finalizado; Cantidad reportada: 27,396 kits; Lote: D100622 Product Code Lot Number Expiration Date 199102S 237115 8/31/2012 199102S 238568 9/30/2012 199102S 238569 11/30/2012 199102S 237493 6/30/2012 2993 235489 11/30/2012 2993 238599 3/31/2013 2993SAMP 238369 3/31/2013 SAMP199102S 237493 6/30/2012 SAMP199102S 283569 11/30/2012. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The lyophilized thrombin component of the Surgiflo Hemostatic Matrix kit with Thrombin has the potential to be non-sterile due to an undetected air leak which occurred during the manufacturing process.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Ethicon, Inc.
Distribuidor
USA (nationwide)
Lote
D100622 Product Code Lot Number Expiration Date 199102S 237115 8/31/2012 199102S 238568 9/30/2012 199102S 238569 11/30/2012 199102S 237493 6/30/2012 2993 235489 11/30/2012 2993 238599 3/31/2013 2993SAMP 238369 3/31/2013 SAMP199102S 237493 6/30/2012 SAMP199102S 283569 11/30/2012
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Ethicon Surgiflo Hemostatic Matrix Kit with Thrombin Plus Flex Tip Distributed by Ethicon Ethicon Inc., Somerville, NJ 08876 USA Manufactured by Ferrosan Medical Devices Ferrosan Medical Devices A/S Sydmarken 5, DK-2860 Soeberg, Denmark Mixed with sterile saline or thrombin solution is indicated for surgical procedures (except opthalmic) for hemostasis when control of capillary, venous and arteriolar bleeding by pressure, ligature, and other conventional procedures is ineffective or impractical. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1897-2012 el 4 de julio de 2012. Peligro reportado: The lyophilized thrombin component of the Surgiflo Hemostatic Matrix kit with Thrombin has the potential to be non-sterile due to an undetected air leak which occurred during the manufacturing process.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter