5.0mm Unit Rod 270mm 5.0mm Unit Rod 290mm 5.0mm Unit Rod 310mm 5.0mm Unit Rod 330mm 5.0mm Unit Rod 350mm 5.0mm Unit Rod 370mm 5.0mm Unit Rod 390mm 5.0mm Unit Rod 410mm 5.0mm Unit Rod 430mm 5.0mm Unit Rod 450mm pedicle screw spinal system; Intended to provide immobilization and stabilization of spinal segments in skeletally mature patients.
Publicado el 21 de agosto de 2013 · Inicio del retiro: 5 de junio de 2013
Resumen
FDA registra el retiro Z-1935-2013 de dispositivos médicos de Clase II, publicado el 21 de agosto de 2013 e iniciado el 5 de junio de 2013. Empresa responsable: Synthes USA HQ, Inc.; Estado: Finalizado; Cantidad reportada: 492; Lote: Part Numbers: 298.269 298.270 298.271 298.272 298.273 298.274 298.275 298.276 298.277 298.278 with lot numbers: 4729951; 4841209; 4923651; 4923652; 4987688; 4987750; 5350635; 2002330; 3000595 4729952; 4841210; 4923653; 4923654; 4987683; 4856248; 4987755; 5153848; 5153853; 1602596; 1880489 4729953; 4841211; 4923666; 4923667; 4987778; 4856275; 4987761; 5159826; 5157329; 3016224 4729954; 4841212; 4919162; 4923669; 4987779; 4856250; 4987766; 5153796; 3093619 4729956; 4923674; 4987825; 4987780; 5066363 4729958; 4923615; 4987826; 4987781; 5066364 4729959; 4841213; 4923655; 4936270; 4987684; 4987771; 5153849; 5153854 4729960; 4835367; 4923656; 4923657; 4984982; 4856246; 4987772; 5159827; 5157330 4729961; 4835368; 4923658; 4923659; 4987685; 4856249; 4987773; 1602617; 3080451 4729962; 4835378; 4919165; 4923660; 4987686; 4856252; 4987776; 5153851. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
DePuy Synthes has issued a Medical Device Correction notice for the 5.0mm Stainless Steel rods. This product was produced using a finishing process not identified as part of the manufacturing specification. No adverse event reports associated with the use of this product were received, and an extensive engineering and medical safety review concluded that the product poses no incremental risk to t
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Synthes USA HQ, Inc.
- Distribuidor
- Nationwide Distribution.
- Lote
- Part Numbers: 298.269 298.270 298.271 298.272 298.273 298.274 298.275 298.276 298.277 298.278 with lot numbers: 4729951; 4841209; 4923651; 4923652; 4987688; 4987750; 5350635; 2002330; 3000595 4729952; 4841210; 4923653; 4923654; 4987683; 4856248; 4987755; 5153848; 5153853; 1602596; 1880489 4729953; 4841211; 4923666; 4923667; 4987778; 4856275; 4987761; 5159826; 5157329; 3016224 4729954; 4841212; 4919162; 4923669; 4987779; 4856250; 4987766; 5153796; 3093619 4729956; 4923674; 4987825; 4987780; 5066363 4729958; 4923615; 4987826; 4987781; 5066364 4729959; 4841213; 4923655; 4936270; 4987684; 4987771; 5153849; 5153854 4729960; 4835367; 4923656; 4923657; 4984982; 4856246; 4987772; 5159827; 5157330 4729961; 4835368; 4923658; 4923659; 4987685; 4856249; 4987773; 1602617; 3080451 4729962; 4835378; 4919165; 4923660; 4987686; 4856252; 4987776; 5153851
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDASynthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicated8 de enero de 2014
- FDASynthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedge18 de diciembre de 2013
- FDASynthes Trochanteric Fixation Nail As part of the Synthes Trochanteric Fixation Nail (TFN) System, the TFN18 de diciembre de 2013
- FDASynthes Soft Tissue Retractor Small Extendible. Offset blade facilitates easy preparation of the epipereosteal11 de diciembre de 2013
- FDATray Surgical, ENT (various instrument trays) Used for stot age and transportation of reusable surgical21 de agosto de 2013
- FDAC-QUR Mesh (All Sizes, shapes). Intended for use in soft tissue deficiencies.21 de agosto de 2013
Preguntas sobre este retiro
¿5.0mm Unit Rod 270mm 5.0mm Unit Rod 290mm 5.0mm Unit Rod 310mm 5.0mm Unit Rod 330mm 5.0mm Unit Rod 350mm 5.0mm Unit Rod 370mm 5.0mm Unit Rod 390mm 5.0mm Unit Rod 410mm 5.0mm Unit Rod 430mm 5.0mm Unit Rod 450mm pedicle screw spinal system; Intended to provide immobilization and stabilization of spinal segments in skeletally mature patients. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1935-2013 el 21 de agosto de 2013. Peligro reportado: DePuy Synthes has issued a Medical Device Correction notice for the 5.0mm Stainless Steel rods. This product was produced using a finishing process not identified as part of the manufacturing specification. No adverse event reports associated with the use of this product were received, and an extensive engineering and medical safety review concluded that the product poses no incremental risk to t
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter