GYRUS ACMI, REF 3740, Everest BiCOAG Bipolar Dissecting Forceps 5mm/45 cm, Sterile EO. GYRUS ACMI, REF 3740PK, or REF 94201PK, PKS LYONS, Dissecting Forceps 5mm/45cm, Sterile EO. The Dissectors are sterile, single use electrosurgical devices that are indicated for the electrosurgical coagulation, mechanical grasping, and dissection of tissue during the performance of laparoscopic and general surgical procedures.
Publicado el 28 de agosto de 2013 · Inicio del retiro: 24 de julio de 2013
Resumen
FDA registra el retiro Z-1983-2013 de dispositivos médicos de Clase II, publicado el 28 de agosto de 2013 e iniciado el 24 de julio de 2013. Empresa responsable: Gyrus Medical, Inc; Estado: Finalizado; Cantidad reportada: 1140; Lote: REF 3740 150080AC 258624MC G166568CC 381809KD 408020AE JF480565 NY492562 REF 3740PK 128971JB 226730IC 226742IC 226743IC 431894CE 431895CE G128677JB REF 94201PK 102290EB 147428AC 167021CC 198596GC 204439HC 204451HC 249022KC 278142BD 320659DD 344712FD 363953ID 373761JD 389095MD 418231BE G386412KD G461037EE JF493195. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
On April 29, 2013 Olympus Surgical Technologies America R&D received results from an accelerated aging study conducted as part of a new market submission. The study found that one of 60 Dissector units that underwent 3 year accelerated aging failed the dye penetration test indicating that the seal on sterile packaging was compromised. The dissector products are labeled with a 3 year shelf life.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Gyrus Medical, Inc
- Distribuidor
- Worldwide Distribution - USA (nationwide) including the states of AL, AZ, CA, FL, GA, IL, IN, KS,KY, MD, MA, MI, MN, MS, MO, NY, NC, OH, PA, SD, TN, TX, VA and WI., and the countries of BRAZIL, CHINA, AUSTRALIA, CHINA, GERMANY and INDIA.
- Lote
- REF 3740 150080AC 258624MC G166568CC 381809KD 408020AE JF480565 NY492562 REF 3740PK 128971JB 226730IC 226742IC 226743IC 431894CE 431895CE G128677JB REF 94201PK 102290EB 147428AC 167021CC 198596GC 204439HC 204451HC 249022KC 278142BD 320659DD 344712FD 363953ID 373761JD 389095MD 418231BE G386412KD G461037EE JF493195
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAGYRUS ACMI, REF 3345 -Everest BiCOAG MOLly Forceps 5mm/45cm REF 3345PK - PKS MOLly Forceps 5mm/45cm REF 364029 de octubre de 2014
- FDAGyrus ACMI Falope-Ring Dilator, item 000878-4. The Falope-Ring Dilator is a conical plastic device that is a27 de agosto de 2014
- FDAGyrus ACMI Disposable Falope-Ring Band Applicator Kits. REF 006889-901 and REF 005280-901 (with 8 mm19 de marzo de 2014
- FDAEndoGastric Solutions EsophyX2 Device with Serosa Fastener and Accessories. Models R2001 and R2002. Indicated28 de agosto de 2013
- FDAProduct Name: nanoPOINT Injector System, item number 37-0000-08. Intended to fold and insert a STAAR Surgical28 de agosto de 2013
- FDAMoore Hip Prosthesis Long Fenestrated Stem, several head Diameter, 203 mm Stem length. Used to seat the28 de agosto de 2013
Preguntas sobre este retiro
¿GYRUS ACMI, REF 3740, Everest BiCOAG Bipolar Dissecting Forceps 5mm/45 cm, Sterile EO. GYRUS ACMI, REF 3740PK, or REF 94201PK, PKS LYONS, Dissecting Forceps 5mm/45cm, Sterile EO. The Dissectors are sterile, single use electrosurgical devices that are indicated for the electrosurgical coagulation, mechanical grasping, and dissection of tissue during the performance of laparoscopic and general surgical procedures. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1983-2013 el 28 de agosto de 2013. Peligro reportado: On April 29, 2013 Olympus Surgical Technologies America R&D received results from an accelerated aging study conducted as part of a new market submission. The study found that one of 60 Dissector units that underwent 3 year accelerated aging failed the dye penetration test indicating that the seal on sterile packaging was compromised. The dissector products are labeled with a 3 year shelf life.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter