Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatographic method for detection of human hemoglobin in fecal samples.
Publicado el 28 de agosto de 2013 · Inicio del retiro: 17 de julio de 2013
Resumen
FDA registra el retiro Z-1997-2013 de dispositivos médicos de Clase II, publicado el 28 de agosto de 2013 e iniciado el 17 de julio de 2013. Empresa responsable: Beckman Coulter Inc.; Estado: Finalizado; Cantidad reportada: 68,116 units; Lote: Part Number 395067. All In-date lots (42 lots total). Lot Number: Manuf. Date: Exp. Date: M207143 27Jul2012 01/31/2014 M208454 21Sep2012 02/28/2014 M210079 19Oct2012 03/31/2014 M211100 30Nov2012 05/31/2014 M212122 20Dec2012 06/30/2014 M212203 17Jan2013 06/30/2014 M301418 22Feb2013 & 28Feb2013 07/31/2014 M202086 24Feb2012 07/31/2013 M203081 16Mar2012 07/31/2013 M203081A 23Mar2012 07/31/2013 M203262 06Apr2012 08/31/2013 M203098 13Apr2012 08/31/2013 M203098A 20Apr2012 08/31/2013 M203098B 25Apr2012 08/31/2013 M203080 27Apr2012 10/31/2013 M203080A 04May2012 10/31/2013 M203080B 11May2012 10/31/2013 M204276 18May2012 09/30/2013 M205199 12May2012 09/30/2013 M205288 22May2012 11/30/2013 M206021 06June2012 11/30/2013 M206166 18June2012 12/31/2013 M207072 05July2012 12/31/2013 M207143 27July2012 01/31/2014 M206259 26June2012 12/31/2013 M207332 30July2012 01/31/2014 M208345 22Aug2012 01/31/2014 M208272 24Aug2012 02/28/2014 M208454 06Sep2012 02/28/2014 M209135 19Sep2012 03/31/2014 M210079 04Oct2012 03/31/2014 M210169 27Oct2012 04/30/2014 M211100 10Nov2012 05/31/2014 M211078 01Dec2012 05/31/2014 M212122 12Dec2012 06/30/2014 M212203 03Jan2013 06/30/2014 M301324 16Jan2013 07/31/2014 M301418 11Feb2013 07/31/2014 M302410 27Feb2013 08/31/2014 M303157 15Mar2013 09/30/2014 M303223 26Mar2013 09/30/2014 M304130 09Apr2013 09/30/2014. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Beckman Coulter has received a customer complaint that suggests trending of positive results has increased from approximately 4% to 1 0%. As a result of our investigation, Beckman Coulter has confirmed an increase in sensitivity at the borderline positive level. " This testing also demonstrated that the issue does not affect results reported as negative.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Beckman Coulter Inc.
- Distribuidor
- Worldwide Distribution - USA and the country of Canada.
- Lote
- Part Number 395067. All In-date lots (42 lots total). Lot Number: Manuf. Date: Exp. Date: M207143 27Jul2012 01/31/2014 M208454 21Sep2012 02/28/2014 M210079 19Oct2012 03/31/2014 M211100 30Nov2012 05/31/2014 M212122 20Dec2012 06/30/2014 M212203 17Jan2013 06/30/2014 M301418 22Feb2013 & 28Feb2013 07/31/2014 M202086 24Feb2012 07/31/2013 M203081 16Mar2012 07/31/2013 M203081A 23Mar2012 07/31/2013 M203262 06Apr2012 08/31/2013 M203098 13Apr2012 08/31/2013 M203098A 20Apr2012 08/31/2013 M203098B 25Apr2012 08/31/2013 M203080 27Apr2012 10/31/2013 M203080A 04May2012 10/31/2013 M203080B 11May2012 10/31/2013 M204276 18May2012 09/30/2013 M205199 12May2012 09/30/2013 M205288 22May2012 11/30/2013 M206021 06June2012 11/30/2013 M206166 18June2012 12/31/2013 M207072 05July2012 12/31/2013 M207143 27July2012 01/31/2014 M206259 26June2012 12/31/2013 M207332 30July2012 01/31/2014 M208345 22Aug2012 01/31/2014 M208272 24Aug2012 02/28/2014 M208454 06Sep2012 02/28/2014 M209135 19Sep2012 03/31/2014 M210079 04Oct2012 03/31/2014 M210169 27Oct2012 04/30/2014 M211100 10Nov2012 05/31/2014 M211078 01Dec2012 05/31/2014 M212122 12Dec2012 06/30/2014 M212203 03Jan2013 06/30/2014 M301324 16Jan2013 07/31/2014 M301418 11Feb2013 07/31/2014 M302410 27Feb2013 08/31/2014 M303157 15Mar2013 09/30/2014 M303223 26Mar2013 09/30/2014 M304130 09Apr2013 09/30/2014
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDADxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C6352014 de octubre de 2026
- FDADxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AU14 de octubre de 2026
- FDABeckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module26 de agosto de 2026
- FDABeckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE,26 de agosto de 2026
- FDAGENESIS(R) II CRUCIATE RETAINING ARTICULAR INSERT, TRIAL, SIZE 3-4, A/P 48 MM, M/L 68 MM, QTY 1, NON-STERILE,28 de agosto de 2013
- FDA00597001101 Femur CR PRECOAT FEM COMP SIZE AML, Rx, Sterile; 00597001102 Femur CR PRECOAT FEM COMP SIZE AMR,28 de agosto de 2013
Preguntas sobre este retiro
¿Hemoccult ICT (Immunochemical Fecal Occult Blood Test) is a rapid, visually read, qualitative immunochemical chromatographic method for detection of human hemoglobin in fecal samples. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1997-2013 el 28 de agosto de 2013. Peligro reportado: Beckman Coulter has received a customer complaint that suggests trending of positive results has increased from approximately 4% to 1 0%. As a result of our investigation, Beckman Coulter has confirmed an increase in sensitivity at the borderline positive level. " This testing also demonstrated that the issue does not affect results reported as negative.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter