Heraeus, PALAMIX duo. Material Number: 66057897.

Publicado el 6 de mayo de 2026 · Inicio del retiro: 25 de marzo de 2026

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Heraeus Medical GmbH (Dental Division)
Distribuidor
Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;
Lote
Material Number: 66057897; UDI-DI: 04260102133617; Batch: all batches with numbers below 520215;
Clasificación de riesgo
Class III
Área (FDA)
device
País
Germany