Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan
Publicado el 10 de mayo de 2017 · Inicio del retiro: 23 de marzo de 2017
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of salt deposits on unprotected electronic components may combust due to heat generated by electronics when instrument cycles. Completed CAPA identified the root cause as supplier providing defective component (sensor ready valve) and the misalignment of the electronic protective shield during manufacturing of the device.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Sysmex America Inc
- Distribuidor
- Nationwide, Canada
- Lote
- **Model: XN-10, Product Number: AP795756, Serial Numbers: 11001 29656, 29668-29786, 29788-29798, 29800-29801, 29803, 29804, 29807, 29812, 29816-29867 **Model: XN-11, Product Number: CS297931, Serial Numbers: A1001-A1124, A1127, A1130, A1135-A1140, A1142-A1144, A1146-A1149, A1151, A1153-A1159, A1161, A1163, A1171
- Clasificación de riesgo
- Class II