00 4309-049-00 Trabecular Metal Reverse Shoulder Glenosphere Distractor Rx, Non sterile (sterilize before use) The Glenosphere Distractor must be able to attach, connect, or mate with the Glenosphere and remove it without dislodging or loosening the base plate from the glenoid, or damaging the glenoid. Pulling the trigger releases a weight inside the Distractor. This force is transmitted to the bottom of the glenosphere dislodging it from the base plate.
Publicado el 4 de septiembre de 2013 · Inicio del retiro: 1 de agosto de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Zimmer is initiating a lot specific recall of the Trabecular Metal" Reverse Glenosphere Distractor due to the potential of the device not functioning properly should it be utilized without proper lubrication. As a result, there is a potential for the device to fail to provide impact to the Glenosphere when the trigger is pulled. In addition, there is a potential for the foot of the device to fract
Qué hacer
Voluntary: Firm initiated
Detalles del retiro
- Fabricante
- Zimmer, Inc.
- Distribuidor
- Worldwide Distribution - USA (nationwide) including the countries of Singapore, Brazil, Taiwan, Australia, Canada, Germany(with further distribution to UAE, Austria, Belgium, Switzerland, Czech Republic, Denmark, Spain, Finland, France, Greece, Israel, Italy, Malta, Norway, Portugal, Sweden and South Africa,
- Lote
- Item # 00-4309-049-00 and Lot Numbers: 60452884, 60461858, 60531990, 60549559, 60605842, 60612781, 60684768, 60754281, 60815778, 60909547, 60959086, 60978338, 61029048, 61135953, 61213252, 61274877, 61334929, 61431672, 61566464, 61594740, 61690097, 61756058, 61857748, 61858851, 61889174, 61906350, 61910781, 61918666 62039450, 62096379, 62101249, 62141887, 62160965, 62215378, 62254111, 62269856, 62304206, and 62311096.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States