TL-HEX TruLok Hexapod System, Double Row Footplates, TL-HEX Sterile, All Sizes: 120 MM, 140 MM, 160 MM, 180 MM, 200 MM. The TL-HEX TrueLock Hexapod System (TL-HEX) consists of rings, foot plates and struts to be used in conjunction with the TruLok external fixation system. The system is a metal bone fixation device for stabilization and correction during the normal healing process. The purpose of the TL-HEX System is to provide a solution for deformity correction and keeping the bone segments aligned during the healing process. The device may only be applied by a physician, who is fully responsible for the surgical and post-operative procedures of the system and is able to manage possible mechanical and software limitations. Indications for use for adults and all pediatric subgroups except newborns include: post-traumatic joint contracture which has resulted in loss of range of motion, fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction, open and closed fracture fixation, pseudoarthrosis of long bones, limb lengthening by epiphyseal or metaphyseal distraction, correction of bony or soft tissue deformities, correction of bony of soft tissue defects, joint arthrodesis, infected fractures or non unions.
Publicado el 13 de junio de 2018 · Inicio del retiro: 30 de enero de 2018
Resumen
FDA registra el retiro Z-2063-2018 de dispositivos médicos de Clase II, publicado el 13 de junio de 2018 e iniciado el 30 de enero de 2018. Empresa responsable: Orthofix Srl; Estado: Finalizado; Cantidad reportada: 408 units; Lote: a) 120 MM, Product Code 99-56-22000, Lot No. V1374370 (Lot No. laser marked on products V1369866, V1369865, V1369864), Exp. Date 06/08/2019, UDI 18054242512585; b) 140 MM, Product Code 99-56-22020, Lot No. V1374298 (Lot No. laser marked on products V1369870, V1369868, V1369869, V1369872, V1369871), Exp. Date 06/07/2019, UDI 18054242512592; c) 160 MM, Product Code 99-56-22040, Lot No. V1374255 (Lot No. laser marked on products V1370134, V1370136, V1370128, V1370133, V1370132, V1370131), Exp. Date 06/07/2019, Lot No. V1374374 (Lot No. laser marked on products V1370130), Exp. Date 06/08/2019, Lot No. R1065030 (Lot No. laser marked on products V1370134, V1370136, V1370128, V1370133, V1370132, V1370131), Exp. Date 12/16/2021, UDI 18054242512608; d) 180 MM, Product Code 99-56-22060, Lot No. V1374259 (Lot No. laser marked on products V1370581, V1369468, V1369469, V1369467, V1369466, V1369465, V1369471, V1369470, V1370597), Exp. Date 06/07/2019, UDI 18054242512615; e) 200 MM, Product Code 99-56-22080, Lot No. V1374361 (Lot No. laser marked on products V1369462, V1371128, V1371074, V1369463, V1370469), Exp. Date 06/08/2019, UDI 18054242512622; f) 220 MM, Product Code 99-56-22100, Lot No. V1374373 (Lot No. laser marked on products V1369460, V1371250, V1371075), Exp. Date 06/08/2019, UDI 18054242512639. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The firm received complaints regarding the incompatibility and dimensional interference between the double row footplates manufactured in a limited period of 2014 and struts manufactured after September 2016. The firm determined that certain batches of the double row footplates are not compatible with the current version of the TL-HEX struts.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Orthofix Srl
- Distribuidor
- Worldwide Distribution - US Distribution within Texas., and to the countries of : Australia, Belgium, Chile, Denmark, France, Germany, Israel, Italy, Japan, South Africa, Spain, Sweden, Switzerland and United Kingdom.
- Lote
- a) 120 MM, Product Code 99-56-22000, Lot No. V1374370 (Lot No. laser marked on products V1369866, V1369865, V1369864), Exp. Date 06/08/2019, UDI 18054242512585; b) 140 MM, Product Code 99-56-22020, Lot No. V1374298 (Lot No. laser marked on products V1369870, V1369868, V1369869, V1369872, V1369871), Exp. Date 06/07/2019, UDI 18054242512592; c) 160 MM, Product Code 99-56-22040, Lot No. V1374255 (Lot No. laser marked on products V1370134, V1370136, V1370128, V1370133, V1370132, V1370131), Exp. Date 06/07/2019, Lot No. V1374374 (Lot No. laser marked on products V1370130), Exp. Date 06/08/2019, Lot No. R1065030 (Lot No. laser marked on products V1370134, V1370136, V1370128, V1370133, V1370132, V1370131), Exp. Date 12/16/2021, UDI 18054242512608; d) 180 MM, Product Code 99-56-22060, Lot No. V1374259 (Lot No. laser marked on products V1370581, V1369468, V1369469, V1369467, V1369466, V1369465, V1369471, V1369470, V1370597), Exp. Date 06/07/2019, UDI 18054242512615; e) 200 MM, Product Code 99-56-22080, Lot No. V1374361 (Lot No. laser marked on products V1369462, V1371128, V1371074, V1369463, V1370469), Exp. Date 06/08/2019, UDI 18054242512622; f) 220 MM, Product Code 99-56-22100, Lot No. V1374373 (Lot No. laser marked on products V1369460, V1371250, V1371075), Exp. Date 06/08/2019, UDI 18054242512639
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Italy
Retiros relacionados
- FDAREF: 99-709245, Fitbone Trochanteric Nail L245mm, D9mm, Sterile25 de junio de 2025
- FDAORTHOFIX Catalogue Number: ref 1100701, drill bit diameter 4.8 MM LENGTH 280MM, RX ONLY, Non Sterile, UDI:20 de noviembre de 2019
- FDAORTHOFIX Catalogue Number: ref 1-1100701,DRILL BIT D.4,8 MM L.280 MM TIN COATED - QUICK CONNECT, RX ONLY, Non20 de noviembre de 2019
- FDAORTHOFIX Catalogue Number: ref: 99-92501, PREFIX II DIAPHYSEAL COMPLETE KIT STERILE SHORT SCREWS, RX ONLY,20 de noviembre de 2019
- FDARX Imola (RX4900) For Professional Use for the quantitative in vitro determination of various clinical13 de junio de 2018
- FDAST-AIA PACK Prolactin; Part Number: 025255 Assay, Reproductive Hormone13 de junio de 2018
Preguntas sobre este retiro
¿TL-HEX TruLok Hexapod System, Double Row Footplates, TL-HEX Sterile, All Sizes: 120 MM, 140 MM, 160 MM, 180 MM, 200 MM. The TL-HEX TrueLock Hexapod System (TL-HEX) consists of rings, foot plates and struts to be used in conjunction with the TruLok external fixation system. The system is a metal bone fixation device for stabilization and correction during the normal healing process. The purpose of the TL-HEX System is to provide a solution for deformity correction and keeping the bone segments aligned during the healing process. The device may only be applied by a physician, who is fully responsible for the surgical and post-operative procedures of the system and is able to manage possible mechanical and software limitations. Indications for use for adults and all pediatric subgroups except newborns include: post-traumatic joint contracture which has resulted in loss of range of motion, fractures and disease which generally may result in joint contractures or loss of range of motion and fractures requiring distraction, open and closed fracture fixation, pseudoarthrosis of long bones, limb lengthening by epiphyseal or metaphyseal distraction, correction of bony or soft tissue deformities, correction of bony of soft tissue defects, joint arthrodesis, infected fractures or non unions. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2063-2018 el 13 de junio de 2018. Peligro reportado: The firm received complaints regarding the incompatibility and dimensional interference between the double row footplates manufactured in a limited period of 2014 and struts manufactured after September 2016. The firm determined that certain batches of the double row footplates are not compatible with the current version of the TL-HEX struts.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter