Rejuvenate Modular Stems Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland Rejuvenate Modular Necks Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430. Stem Bodies Rejuvenate Modular consists of 6 modular stems bodies in sizes 7-12. The Modular Stem bodies are made of TMZF alloy, developed specifically for orthopaedic applications. Rejuvenate Modular Necks The Rejuvenate Modular Necks are made of Cobalt Chrome. The necks in this system consist of 4 lengths in 4 mm increments. They are 30 mm, 34 mm, 38 mm and 42 mm. They come in multiple neck angles and versions in each length. They are 127, 130 and 132. The versions are 0, 8, 16 degrees. In total there are 16 unique neck implants in the Rejuvenate Modular System.
Publicado el 1 de agosto de 2012 · Inicio del retiro: 30 de abril de 2012
Resumen
FDA registra el retiro Z-2090-2012 de dispositivos médicos de Clase II, publicado el 1 de agosto de 2012 e iniciado el 30 de abril de 2012. Empresa responsable: Stryker Howmedica Osteonics Corp.; Estado: En curso; Cantidad reportada: (NLS/NLV) - 23193 units and (SPT) - 20,743 units; Lote: 510 K092561 all lots Rejuvenate (NLS/NLV) Rejuvenate (SPT) (4845-4XXX and NLS/NLV). La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems. This is based on a reported rate of less than one percent for revisions potentially associated with fretting and or corrosion at or about the modular neck junction. On June 29, 2012, Stryker has decided to remove the affected product from the field since there may be an
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Stryker Howmedica Osteonics Corp.
- Distribuidor
- Nationwide Distribution
- Lote
- 510 K092561 all lots Rejuvenate (NLS/NLV) Rejuvenate (SPT) (4845-4XXX and NLS/NLV)
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAHoffmann LRF Safety Clip (Bone Transport Strut), 14 x 12 x 5 mm, REF 4933-0-401, NDC31 de mayo de 2017
- FDASupracondylar Nail T2 SCN 09x300 mm T2 Supracondylar Nail System is intended to provide strong and stable21 de septiembre de 2016
- FDAStryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used to17 de agosto de 2016
- FDAStryker, T2 Guidewire, Ball-Tipped, 3 x 1000mm Product Usage: Intended for the temporary stabilization of bone2 de marzo de 2016
- FDAMetformin Hydrochloride Tablets, USP, 1000 mg, Rx only, 1000 count tablets per Bottle, Manufactured by: Cadila8 de agosto de 2012
- FDAABG II Modular Hip Stem Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430 Stryker Ireland Carrigtwohill1 de agosto de 2012
Preguntas sobre este retiro
¿Rejuvenate Modular Stems Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland Rejuvenate Modular Necks Stryker Howmedica Osteonics Corp. Mahwah, NJ 07430. Stem Bodies Rejuvenate Modular consists of 6 modular stems bodies in sizes 7-12. The Modular Stem bodies are made of TMZF alloy, developed specifically for orthopaedic applications. Rejuvenate Modular Necks The Rejuvenate Modular Necks are made of Cobalt Chrome. The necks in this system consist of 4 lengths in 4 mm increments. They are 30 mm, 34 mm, 38 mm and 42 mm. They come in multiple neck angles and versions in each length. They are 127, 130 and 132. The versions are 0, 8, 16 degrees. In total there are 16 unique neck implants in the Rejuvenate Modular System. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2090-2012 el 1 de agosto de 2012. Peligro reportado: Stryker has updated the instruction for use (IFU) (product correction)for the ABGII Modular and Rejuvenate Modular Hip Systems. This is based on a reported rate of less than one percent for revisions potentially associated with fretting and or corrosion at or about the modular neck junction. On June 29, 2012, Stryker has decided to remove the affected product from the field since there may be an
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter