Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2390-2016

Stryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used to complement or replace the standard instruments used during the implantation of hip, knee, trauma and upper extremities implants

Publicado el 17 de agosto de 2016 · Inicio del retiro: 23 de junio de 2016

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

It was reported that the secondary locking mechanism, and its corresponding components, disassociated from the Specialty Triathlon Tibial Alignment Handle with Secondary Lock Assembly during surgery. Upon further investigation, it was discovered that the weld, which was intended to hold the secondary locking mechanism together, did not meet the weld size specified on the engineering drawing.

Qué hacer

Voluntary: Firm initiated · E-Mail

Detalles del retiro

Fabricante
Stryker Howmedica Osteonics Corp.
Distribuidor
Nationwide Distribution to FL, IN, LA, NC, PA, TN and UT
Lote
Catalog No: I-K3254TA00 Lot Nos: F6E13243, F6L13864, F6M14078, F6S14667, F6S14831, F6W15444, F6W15406, F7C15943, F7H16099 and F7L16994
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta