Stryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used to complement or replace the standard instruments used during the implantation of hip, knee, trauma and upper extremities implants
Publicado el 17 de agosto de 2016 · Inicio del retiro: 23 de junio de 2016
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
It was reported that the secondary locking mechanism, and its corresponding components, disassociated from the Specialty Triathlon Tibial Alignment Handle with Secondary Lock Assembly during surgery. Upon further investigation, it was discovered that the weld, which was intended to hold the secondary locking mechanism together, did not meet the weld size specified on the engineering drawing.
Qué hacer
Voluntary: Firm initiated · E-Mail
Detalles del retiro
- Fabricante
- Stryker Howmedica Osteonics Corp.
- Distribuidor
- Nationwide Distribution to FL, IN, LA, NC, PA, TN and UT
- Lote
- Catalog No: I-K3254TA00 Lot Nos: F6E13243, F6L13864, F6M14078, F6S14667, F6S14831, F6W15444, F6W15406, F7C15943, F7H16099 and F7L16994
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States