Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2113-2013

DeRoyal Insufflation Tubing sets: 1) REF 28-0206 Insufflation Tubing, Laparoscopic Rx Only 2) REF 28-0214 Hi-Flow Insufflation Tubing Rx Only 3) REF 28-0216 Tubing, Insuf w/.3 Micr Filter Rx Only 4) REF 28-0217 Insufflation Tubing, Laparoscopic Rx Only 5) REF 28-0208 Insufflation Tubing w/Vent Rx Only 6) REF 28-0212H Heated Hi-Flow Insufflation Tubing Rx Only. Used during laparoscopic surgical procedures to aid in the insufflation of the abdomen.

Publicado el 4 de septiembre de 2013 · Inicio del retiro: 15 de julio de 2013

Resumen

FDA registra el retiro Z-2113-2013 de dispositivos médicos de Clase II, publicado el 4 de septiembre de 2013 e iniciado el 15 de julio de 2013. Empresa responsable: DeRoyal Industries Inc; Estado: Finalizado; Cantidad reportada: 506,632 sets in US; 17,854 sets International; Lote: All Lots of the Insufflation Tubing Sets are as follows: REF 28-0206 Insufflation Tubing, Laparoscopic, Lot #29074214; REF 28-0214 Hi-Flow Insufflation Tubing Lot #3194170; REF 28-0216 Tubing, Insuf w/.3 Micr Filter Lot #29074249; REF 28-0217 Insufflation Tubing, Laparoscopic Lot #32305010; DeRoyal REF 28-0208 Insufflation Tubing w/Vent Lot #29314302; DeRoyal REF 28-0212H Heated Hi-Flow Insufflation Tubing Lot #29000354. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

DeRoyal Industries is recalling their product Insufflation Tubing Sets due to materials incompatibility which could possibly result in occlusion at or near the filter housing which could prevent the flow of gas through the device.

Qué hacer

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
DeRoyal Industries Inc
Distribuidor
Worldwide Distribution-USA (nationwide) including DC and the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and the countries of Dubai, Arab Emirates, Quebec, Canada, Japan, and Ireland.
Lote
All Lots of the Insufflation Tubing Sets are as follows: REF 28-0206 Insufflation Tubing, Laparoscopic, Lot #29074214; REF 28-0214 Hi-Flow Insufflation Tubing Lot #3194170; REF 28-0216 Tubing, Insuf w/.3 Micr Filter Lot #29074249; REF 28-0217 Insufflation Tubing, Laparoscopic Lot #32305010; DeRoyal REF 28-0208 Insufflation Tubing w/Vent Lot #29314302; DeRoyal REF 28-0212H Heated Hi-Flow Insufflation Tubing Lot #29000354.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿DeRoyal Insufflation Tubing sets: 1) REF 28-0206 Insufflation Tubing, Laparoscopic Rx Only 2) REF 28-0214 Hi-Flow Insufflation Tubing Rx Only 3) REF 28-0216 Tubing, Insuf w/.3 Micr Filter Rx Only 4) REF 28-0217 Insufflation Tubing, Laparoscopic Rx Only 5) REF 28-0208 Insufflation Tubing w/Vent Rx Only 6) REF 28-0212H Heated Hi-Flow Insufflation Tubing Rx Only. Used during laparoscopic surgical procedures to aid in the insufflation of the abdomen. fue retirado del mercado?

Sí. FDA publicó el retiro Z-2113-2013 el 4 de septiembre de 2013. Peligro reportado: DeRoyal Industries is recalling their product Insufflation Tubing Sets due to materials incompatibility which could possibly result in occlusion at or near the filter housing which could prevent the flow of gas through the device.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit