MONOCRYL Plus Antibacterial (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated. PRODUCT CODE: MCP266H13 MCP3200H16 MCP3212H16 MCP3213H16 MCP340H13 MCP345H13 MCP416H14 MCP417H14 MCP426H13 MCP427H13 MCP4423H16 MCP4424H14 MCP442H14 MCP496H16 MCP497H16 MCP604H13 MCP936H13 MCP945H14 MCPB346H13
Publicado el 26 de junio de 2024 · Inicio del retiro: 10 de mayo de 2024
Resumen
FDA registra el retiro Z-2149-2024 de dispositivos médicos de Clase II, publicado el 26 de junio de 2024 e iniciado el 10 de mayo de 2024. Empresa responsable: Ethicon, Inc.; Estado: En curso; Cantidad reportada: 287892 units; Lote: GTIN: 10705031046337 10705031121393 10705031121423 10705031121430 10705031046405 10705031046429 10705031046481 10705031046498 10705031046504 10705031046511 10705031121997 10705031148116 10705031122000 10705031122314 10705031122338 10705031046634 10705031046757 10705031046795 10705031046863 Product Code Lot Code MCP266H UAMMCX MCP340H UAMQUZ MCP416H UBMHLB MCP417H UBMLKK MCP426H UAMLDQ UAMRSL UAMRUM UAMRUU UAMRMR UAMRUK UAMRMB MCP427H UAMLRK UAMMCH UAMLTR UAMKTS UAMKQL UAMREE MCP604H UAMLQE MCP936H UBMDQM UBMDPZ UBMEXS UBMQAR UBMEQX UBMKZA UBMLDR MCP945H UAMRKQ MCPB346H UAMQZB. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Ethicon, Inc.
- Distribuidor
- Worldwide distribution - US Nationwide including PR and the countries of AR, AU, AU, BE, BR, CA, CA, CL, CN, CO, CR, EC, IL, JP, JP, KW, MX, NZ, PA, PR, PR, SA, SG, UY, ZA.
- Lote
- GTIN: 10705031046337 10705031121393 10705031121423 10705031121430 10705031046405 10705031046429 10705031046481 10705031046498 10705031046504 10705031046511 10705031121997 10705031148116 10705031122000 10705031122314 10705031122338 10705031046634 10705031046757 10705031046795 10705031046863 Product Code Lot Code MCP266H UAMMCX MCP340H UAMQUZ MCP416H UBMHLB MCP417H UBMLKK MCP426H UAMLDQ UAMRSL UAMRUM UAMRUU UAMRMR UAMRUK UAMRMB MCP427H UAMLRK UAMMCH UAMLTR UAMKTS UAMKQL UAMREE MCP604H UAMLQE MCP936H UBMDQM UBMDPZ UBMEXS UBMQAR UBMEQX UBMKZA UBMLDR MCP945H UAMRKQ MCPB346H UAMQZB
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACoated VICRYL Polyglactin 910 Suture VIO 36IN(90CM) USP0(M3.5) S/A CT, Product Number J358H13 de noviembre de 2024
- FDAVICRYL (polyglactin 910) Suture - Indicated for use in general soft tissue approximation and/or ligation,26 de junio de 2024
- FDAPDS Plus Antibacterial (polydioxanone) Suture - Intended for use in general soft tissue approximation,26 de junio de 2024
- FDAMONOCRYL (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation26 de junio de 2024
- FDATabloCart with Prefiltration Drawer PN-0006813, with power cords: PN-0005746 (10-foot), PN-0006578 (15-foot),26 de junio de 2024
- FDACentrella Max - The Centrella Smart+ Bed is intended for use in healthcare environments as a patient support26 de junio de 2024
Preguntas sobre este retiro
¿MONOCRYL Plus Antibacterial (poliglecaprone 25) Suture - Intended for use in general soft tissue approximation and/or ligation where an absorbable material is indicated. PRODUCT CODE: MCP266H13 MCP3200H16 MCP3212H16 MCP3213H16 MCP340H13 MCP345H13 MCP416H14 MCP417H14 MCP426H13 MCP427H13 MCP4423H16 MCP4424H14 MCP442H14 MCP496H16 MCP497H16 MCP604H13 MCP936H13 MCP945H14 MCPB346H13 fue retirado del mercado?
Sí. FDA publicó el retiro Z-2149-2024 el 26 de junio de 2024. Peligro reportado: Issue on a specific packaging machine resulted in a hole in the primary packaging, product sterility compromised and could result in infection .
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter