Tec 6 continuous Flow Vaporizer and Tec 6 Plus Anesthesia Vaporizer. The Datex-Ohmeda Tec 6 Plus and Tec 6 Plus NAD Variant is an electronic vaporizer which delivers the anesthetic agent desflurane. The Tec 6 Plus attaches only to Datex-Ohmeda anesthesia machines equipped with the proprietary Selectatec manifold. The Tec 6 Plus NAD Variant attaches only to NAD anesthesia machines equipped with the NAD exclusion type manifold.
Publicado el 18 de septiembre de 2013 · Inicio del retiro: 29 de julio de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
GE Healthcare has recently become aware of a potential safety issue involving the pre-operative check, low pressure leak test at 1% dial setting of the Tec 6 and Tec 6 Plus Vaporizers. The user manual pre-operative check, low pressure leak test may not detect the full range of leaks from seal wear degradation in the vaporizers. If the leak is not detected, it will result in fresh gas leaking to
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- GE Healthcare, LLC
- Distribuidor
- Worldwide Distribution - USA (nationwide) Puerto Rico and internationally to AUSTRALIA, AUSTRIA, BAHAMAS, BAHRAIN, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CROATIA local name-Hrvatska, CYPRUS, DENMARK, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JORDAN, KOREA-REPUBLIC OF, KUWAIT, LITHUANIA, MALAYSIA, MALTA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PAKISTAN, POLAND, PORTUGAL, QATAR, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, and VENEZUELA,
- Lote
- Part Numbers: X1107-9001-000, X1107-9002-000, X1107-9003-000, X1107-9004-000, X1107-9005-000, X1107-9006-000, X1107-9007-000, X1109-9008-000, X1109-9009-000, X1107-9010-000, X1107-9011-000, X1107-9012-000, X1107-9014-000, X1107-9015-000, X1107-9016-000, X1107-9101-000, X1107-9102-000, X1107-9103-000, X1107-9201-000, X1107-9202-000, X1107-9203-000, X1107-9205-000, X1107-9206-000, X1107-9601-000, X1107-9602-000, X1107-9603-000, X1107-9604-000, X1107-9605-000, X1107-9606-000, X1107-9607-000, X1107-9608-000, X1107-9609-000, X1107-9610-000, X1107-9611-000, X1107-9612-000, X1107-9614-000, X1107-9615-000, X1107-9616-000, X1107-9617-000, M1145610-X, M1220023-X, M1220854-X, M1221495-X, 2064455-X, M1243207-X, X1107-9701-000, X1107-9702-000, X1107-9703-000, X1107-9211-000, X1107-9212-000, X1107-9213-000, X1107-9215-000, X1107-9216-000, X1107-9217-000, X1107-9207-000, X1107-9208-000, 2066725-X, 2066835-X, 2069767-X, 1107-9601-000, 1107-9612-000, M1221495, M1220023, M1243207, 1107-9602-000, 1107-9603-000, 1107-9604-000, 1107-9605-000, 1107-9606-000, 1107-9607-000, 1107-9608-000, 1107-9609-000, 1107-9610-000, 1107-9611-000, 1107-9614-000, 1107-9615-000, 1107-9616-000, 1107-9617-000, M1145610, M1220854, 2064455-001, M1098876, 2066725-001, 2066835-001, 2069767-001, X1107-8003-000, X1107-8006-000, X1107-8007-000, X1107-8008-000, X1107-8012-000, X1107-8013-000, X1107-8014-000, X1107-8015-000, X1107-8016-000, X1107-8017-000, X1107-8018-000, X1107-8019-000
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?