Tec 6 continuous Flow Vaporizer and Tec 6 Plus Anesthesia Vaporizer. The Datex-Ohmeda Tec 6 Plus and Tec 6 Plus NAD Variant is an electronic vaporizer which delivers the anesthetic agent desflurane. The Tec 6 Plus attaches only to Datex-Ohmeda anesthesia machines equipped with the proprietary Selectatec manifold. The Tec 6 Plus NAD Variant attaches only to NAD anesthesia machines equipped with the NAD exclusion type manifold.
Publicado el 18 de septiembre de 2013 · Inicio del retiro: 29 de julio de 2013
Resumen
FDA registra el retiro Z-2161-2013 de dispositivos médicos de Clase II, publicado el 18 de septiembre de 2013 e iniciado el 29 de julio de 2013. Empresa responsable: GE Healthcare, LLC; Estado: Finalizado; Cantidad reportada: 31,443 devices (23,947 US, 7496 OUS); Lote: Part Numbers: X1107-9001-000, X1107-9002-000, X1107-9003-000, X1107-9004-000, X1107-9005-000, X1107-9006-000, X1107-9007-000, X1109-9008-000, X1109-9009-000, X1107-9010-000, X1107-9011-000, X1107-9012-000, X1107-9014-000, X1107-9015-000, X1107-9016-000, X1107-9101-000, X1107-9102-000, X1107-9103-000, X1107-9201-000, X1107-9202-000, X1107-9203-000, X1107-9205-000, X1107-9206-000, X1107-9601-000, X1107-9602-000, X1107-9603-000, X1107-9604-000, X1107-9605-000, X1107-9606-000, X1107-9607-000, X1107-9608-000, X1107-9609-000, X1107-9610-000, X1107-9611-000, X1107-9612-000, X1107-9614-000, X1107-9615-000, X1107-9616-000, X1107-9617-000, M1145610-X, M1220023-X, M1220854-X, M1221495-X, 2064455-X, M1243207-X, X1107-9701-000, X1107-9702-000, X1107-9703-000, X1107-9211-000, X1107-9212-000, X1107-9213-000, X1107-9215-000, X1107-9216-000, X1107-9217-000, X1107-9207-000, X1107-9208-000, 2066725-X, 2066835-X, 2069767-X, 1107-9601-000, 1107-9612-000, M1221495, M1220023, M1243207, 1107-9602-000, 1107-9603-000, 1107-9604-000, 1107-9605-000, 1107-9606-000, 1107-9607-000, 1107-9608-000, 1107-9609-000, 1107-9610-000, 1107-9611-000, 1107-9614-000, 1107-9615-000, 1107-9616-000, 1107-9617-000, M1145610, M1220854, 2064455-001, M1098876, 2066725-001, 2066835-001, 2069767-001, X1107-8003-000, X1107-8006-000, X1107-8007-000, X1107-8008-000, X1107-8012-000, X1107-8013-000, X1107-8014-000, X1107-8015-000, X1107-8016-000, X1107-8017-000, X1107-8018-000, X1107-8019-000. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
GE Healthcare has recently become aware of a potential safety issue involving the pre-operative check, low pressure leak test at 1% dial setting of the Tec 6 and Tec 6 Plus Vaporizers. The user manual pre-operative check, low pressure leak test may not detect the full range of leaks from seal wear degradation in the vaporizers. If the leak is not detected, it will result in fresh gas leaking to
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- GE Healthcare, LLC
- Distribuidor
- Worldwide Distribution - USA (nationwide) Puerto Rico and internationally to AUSTRALIA, AUSTRIA, BAHAMAS, BAHRAIN, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHILE, CHINA, COSTA RICA, CROATIA local name-Hrvatska, CYPRUS, DENMARK, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JORDAN, KOREA-REPUBLIC OF, KUWAIT, LITHUANIA, MALAYSIA, MALTA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PAKISTAN, POLAND, PORTUGAL, QATAR, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWITZERLAND, TAIWAN, THAILAND, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, and VENEZUELA,
- Lote
- Part Numbers: X1107-9001-000, X1107-9002-000, X1107-9003-000, X1107-9004-000, X1107-9005-000, X1107-9006-000, X1107-9007-000, X1109-9008-000, X1109-9009-000, X1107-9010-000, X1107-9011-000, X1107-9012-000, X1107-9014-000, X1107-9015-000, X1107-9016-000, X1107-9101-000, X1107-9102-000, X1107-9103-000, X1107-9201-000, X1107-9202-000, X1107-9203-000, X1107-9205-000, X1107-9206-000, X1107-9601-000, X1107-9602-000, X1107-9603-000, X1107-9604-000, X1107-9605-000, X1107-9606-000, X1107-9607-000, X1107-9608-000, X1107-9609-000, X1107-9610-000, X1107-9611-000, X1107-9612-000, X1107-9614-000, X1107-9615-000, X1107-9616-000, X1107-9617-000, M1145610-X, M1220023-X, M1220854-X, M1221495-X, 2064455-X, M1243207-X, X1107-9701-000, X1107-9702-000, X1107-9703-000, X1107-9211-000, X1107-9212-000, X1107-9213-000, X1107-9215-000, X1107-9216-000, X1107-9217-000, X1107-9207-000, X1107-9208-000, 2066725-X, 2066835-X, 2069767-X, 1107-9601-000, 1107-9612-000, M1221495, M1220023, M1243207, 1107-9602-000, 1107-9603-000, 1107-9604-000, 1107-9605-000, 1107-9606-000, 1107-9607-000, 1107-9608-000, 1107-9609-000, 1107-9610-000, 1107-9611-000, 1107-9614-000, 1107-9615-000, 1107-9616-000, 1107-9617-000, M1145610, M1220854, 2064455-001, M1098876, 2066725-001, 2066835-001, 2069767-001, X1107-8003-000, X1107-8006-000, X1107-8007-000, X1107-8008-000, X1107-8012-000, X1107-8013-000, X1107-8014-000, X1107-8015-000, X1107-8016-000, X1107-8017-000, X1107-8018-000, X1107-8019-000
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACentricity PACS Software Version 7.0 SP0.0.4.73 de agosto de 2022
- FDASIGNA Premier magnetic resonance scanner, model 5748519.27 de julio de 2022
- FDAGE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images and13 de julio de 2022
- FDABack up batteries in Avance CS2 and Avance CS2 Pro anesthesia devices manufactured on or after April 1, 20196 de julio de 2022
- FDAMicroscan rapID/S panels NC3.33 Sold in Japan only. Product Usage: MicroScan® Synergies plus® Panels &18 de septiembre de 2013
- FDATempa DOT Single-Use Clinical Thermometer Medical Indicators Made in U.S.A. by Medical Indicators, Inc. 1618 de septiembre de 2013
Preguntas sobre este retiro
¿Tec 6 continuous Flow Vaporizer and Tec 6 Plus Anesthesia Vaporizer. The Datex-Ohmeda Tec 6 Plus and Tec 6 Plus NAD Variant is an electronic vaporizer which delivers the anesthetic agent desflurane. The Tec 6 Plus attaches only to Datex-Ohmeda anesthesia machines equipped with the proprietary Selectatec manifold. The Tec 6 Plus NAD Variant attaches only to NAD anesthesia machines equipped with the NAD exclusion type manifold. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2161-2013 el 18 de septiembre de 2013. Peligro reportado: GE Healthcare has recently become aware of a potential safety issue involving the pre-operative check, low pressure leak test at 1% dial setting of the Tec 6 and Tec 6 Plus Vaporizers. The user manual pre-operative check, low pressure leak test may not detect the full range of leaks from seal wear degradation in the vaporizers. If the leak is not detected, it will result in fresh gas leaking to
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter