Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, including blood, lipid, and Total Parenteral Nutrition (TPN). Model number: 621-100ES, 621-200ES, 621-300ES, and 621-400ES. To deliver a broad range of drugs with use with the B. Braun Medical Inc. Horizon Pump IV Sets.

Publicado el 18 de septiembre de 2013 · Inicio del retiro: 31 de octubre de 2012

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

This notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on 26 August, 2011 due to an issue in the Outlook ES Safety Infusion System where the Care Area can inadvertently be exited thereby removing DoseGuard dose limit protection. This occurs when a sequence of HOLD - HOLD key strokes are performed on the pump key panel while the user is in the Dose Dat

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
B Braun Medical, Inc.
Distribuidor
Nationwide Distribution.
Lote
Model number: 621-100ES, 621-200ES, 621-300ES, and 621-400ES with main processor software versions prior to 151580. Multiple serial numbers affected. Affected serial numbers were notified of this field correction.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States