Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, including blood, lipid, and Total Parenteral Nutrition (TPN). Model number: 621-100ES, 621-200ES, 621-300ES, and 621-400ES. To deliver a broad range of drugs with use with the B. Braun Medical Inc. Horizon Pump IV Sets.
Publicado el 18 de septiembre de 2013 · Inicio del retiro: 24 de octubre de 2012
Resumen
FDA registra el retiro Z-2163-2013 de dispositivos médicos de Clase II, publicado el 18 de septiembre de 2013 e iniciado el 24 de octubre de 2012. Empresa responsable: B Braun Medical, Inc.; Estado: Finalizado; Cantidad reportada: 28,427; Lote: Multiple serial numbers affected. Model number: 621-100ES, 621-200ES, 621-300ES, and 621-400ES. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
B. Braun Medical Inc. has become aware of an issue with the Outlook ES Safety Infusion System. The key panel may fail to function as intended if the underlying circuitry fails due to intermittent contact. This issue applies specifically to the HOLD button. If this occurs, the Door Processor (DP) will not receive the desired input from the user. This issue can be observed in pumps with key panels
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- B Braun Medical, Inc.
- Distribuidor
- Nationwide Distribution.
- Lote
- Multiple serial numbers affected. Model number: 621-100ES, 621-200ES, 621-300ES, and 621-400ES.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAAPEX Compounding System Control Panel Module, REF: 601237, Product Code AX1000 - Product Usage: intended for30 de diciembre de 2020
- FDAOutlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow25 de septiembre de 2013
- FDAOutlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow18 de septiembre de 2013
- FDAOutlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow18 de septiembre de 2013
- FDATempa DOT Single-Use Clinical Thermometer Medical Indicators Made in U.S.A. by Medical Indicators, Inc. 1618 de septiembre de 2013
- FDATec 6 continuous Flow Vaporizer and Tec 6 Plus Anesthesia Vaporizer. The Datex-Ohmeda Tec 6 Plus and Tec 618 de septiembre de 2013
Preguntas sobre este retiro
¿Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, including blood, lipid, and Total Parenteral Nutrition (TPN). Model number: 621-100ES, 621-200ES, 621-300ES, and 621-400ES. To deliver a broad range of drugs with use with the B. Braun Medical Inc. Horizon Pump IV Sets. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2163-2013 el 18 de septiembre de 2013. Peligro reportado: B. Braun Medical Inc. has become aware of an issue with the Outlook ES Safety Infusion System. The key panel may fail to function as intended if the underlying circuitry fails due to intermittent contact. This issue applies specifically to the HOLD button. If this occurs, the Door Processor (DP) will not receive the desired input from the user. This issue can be observed in pumps with key panels
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter