Liquid Urine Control Level 3 For Professional Use in the quality control monitoring of In-Vitro Diagnostic Urinary Clinical Assays on a range of clinical chemistry systems.
Publicado el 7 de junio de 2017 · Inicio del retiro: 27 de abril de 2017
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
According to Randox Laboratories an investigation indicates a labelling error for 2 of the test methods listed for quantification of Total Protein (urine). The values for the mean of all instruments are incorrect by a factor of 10. Instrument specific values are correct. This does not affect the performance or stability claims of the product. As control results will be outside of the quoted range this may result in a delay in reporting the sample test results. However as the correct instrument specific means are also included in the value sheet this delay is unlikely. Customers have been referred to their Medical Director for further advice
Qué hacer
Voluntary: Firm initiated · E-Mail
Detalles del retiro
- Fabricante
- Randox Laboratories
- Distribuidor
- Worldwide Distribution - US Distribution to Puerto Rico and to the countries of : Canada Kuwait Chile Poland China Romania France Saudi Arabia Guatemala Slovenia Iran UAE India UK Italy Uruguay Iraq
- Lote
- Catalogue number UC5075 Batch number 907UC
- Clasificación de riesgo
- Class III
- Área (FDA)
- device
- País
- United States