i-Stat cTnl cartridges Abbott Point of Care Inc. The i-Stat cTnl test is an in vitro diagnostic test for the quantitative measurement of cardiac troponin I in heparinized whole blood or plasma samples. Cardiac troponin I measurements can be used as an aid in the diagnosis and treatment of myocardial infarction and in the risk stratification of patients with acute coronary syndromes with respect to their relative risk of mortality. The cartridge is to be used with the i-Stat Analyzer, but not with the i-Stat Portable Clinical Analyzer or the Philips Medical System (formerly Agilent Technologies) Blood Analysis Module (BAM). As part of the i-Stat System, the cTnl test is to be used by trained health care professionals in accordance with a facility's policies and procedures.
Publicado el 15 de agosto de 2012 · Inicio del retiro: 1 de febrero de 2012
Resumen
FDA registra el retiro Z-2169-2012 de dispositivos médicos de Clase II, publicado el 15 de agosto de 2012 e iniciado el 1 de febrero de 2012. Empresa responsable: Abbott Point Of Care Inc.; Estado: Finalizado; Cantidad reportada: 7,057,675 cartridges US ; 772,400 OUS; Lote: List number 06F15-03, 06F15-04, 03P90-23, 600-9009-25 All lots. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The frequency of suppressed results for i-Stat cTnl, BNP and CK-MP cartridges is affected by atmospheric pressure.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Abbott Point Of Care Inc.
- Distribuidor
- Worldwide Distribution - USA including AZ and CO and internationally to Argentina, Australia, Austria, Belgium, Bolivia, Brazil, Chile, China, Colombia, Costa Rica, Dominican Republic, El Salvador, Finland, France, Georgia, Germany, Greece, Honduras, Hong Kong, India, Ireland, Italy, Korea, Kuwait, Mauritius, Mexico, Netherland (Antilles), Nigeria, Norway, Oman, Paraguay, Peru, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, R.O.C., Thailand, UAE, United Kingdom, and Yemen.
- Lote
- List number 06F15-03, 06F15-04, 03P90-23, 600-9009-25 All lots
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAi-STAT EG6+ cartridge; List Number: 03P77-25;6 de mayo de 2026
- FDAi-STAT G3+ cartridge; List Number: 03P78-26;6 de mayo de 2026
- FDAi-STAT EG7+ cartridge; List Number: 03P76-25;6 de mayo de 2026
- FDAi-STAT EG6+ cartridge. List Number: 03P77-25.24 de septiembre de 2025
- FDASmith & Nephew 5.5 mm TWINFIX Ultra Ti Anchor with three #2 sutures (white, COBRAID- blue), sterile Part15 de agosto de 2012
- FDAi-Stat BNP cartridges Abbott Point of Care Inc. The i-Stat BNP test is an in vitro diagnostic test for the15 de agosto de 2012
Preguntas sobre este retiro
¿i-Stat cTnl cartridges Abbott Point of Care Inc. The i-Stat cTnl test is an in vitro diagnostic test for the quantitative measurement of cardiac troponin I in heparinized whole blood or plasma samples. Cardiac troponin I measurements can be used as an aid in the diagnosis and treatment of myocardial infarction and in the risk stratification of patients with acute coronary syndromes with respect to their relative risk of mortality. The cartridge is to be used with the i-Stat Analyzer, but not with the i-Stat Portable Clinical Analyzer or the Philips Medical System (formerly Agilent Technologies) Blood Analysis Module (BAM). As part of the i-Stat System, the cTnl test is to be used by trained health care professionals in accordance with a facility's policies and procedures. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2169-2012 el 15 de agosto de 2012. Peligro reportado: The frequency of suppressed results for i-Stat cTnl, BNP and CK-MP cartridges is affected by atmospheric pressure.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter