FDAFDA · Alimentos y medicamentos·N.º oficial Z-2176-2012

Stryker Reliance AL Implant Blocker (Various sizes) The SLIDE Implant Blocker is a reusable instrument, sterilized prior to use. It allows the surgeon to distract the disc space in the lumbar region of the spine using an anterior or anterolateral approach while also allowing them to insert the implant into final position.

Publicado el 15 de agosto de 2012 · Inicio del retiro: 21 de marzo de 2011

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Stryker Spine has received six reports regarding the breakage of one or more prongs at the extremity of the instrument.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Stryker Spine
Distribuidor
Nationwide Distribution including NC, AR, TN, MA, IL, GA, NY, LA, MN, MO, OH, KS, NM, TX, KS,and IN.
Lote
Catalog Number: 48361223, 48361224, 48361225, 48361226, 48361227, 48361228 Lot Code: 084505, 084506, 084507, 084508, 084509, 084510
Clasificación de riesgo
Class II
Área (FDA)
device