Dolphin Inflation Device, REF 0185NA. Rx only, Sterilized using Ethylene Oxide. DOLPHIN is a single-use, sterile inflation device used in cardiovascular procedures to pressurize (inflate) and depressurize (deflate) balloon catheters. The device itself consists of two components - the inflation device and large volume syringe (30 cc). The manually operation of the DOLPHIN is achieved by the manipulation of a large handle to drive a piston housed within the body of the device. Careful and controlled inflation is achieved by rotating the handle clockwise. During inflation a unique cam locking mechanism maintains pressure even if the user lets go of the device. Instantaneous deflation, regardless of balloon size, is made possible by the release of the dual locks located on device as well as a large, 30 cc syringe that is sold with each Dolphin Inflation Device. All the while during inflation or deflation pressure is displayed and can be monitored on a large analog gauge mounted on top of the device.
Publicado el 7 de junio de 2017 · Inicio del retiro: 12 de abril de 2017
Resumen
FDA registra el retiro Z-2180-2017 de dispositivos médicos de Clase II, publicado el 7 de junio de 2017 e iniciado el 12 de abril de 2017. Empresa responsable: Vascular Solutions, Inc.; Estado: Finalizado; Cantidad reportada: 4,110; Lote: Lot Numbers: 15075668, 15075924, 15115426, 15125175, 16025324, 16035493, 16055261. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The manufacturer of the DOLPHIN inflation device, Perouse Medical, has initiated a recall of the device due to a potential defect in the sterile barrier. Use of affected product could pose a risk of microbial contamination, leading to infection.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Vascular Solutions, Inc.
- Distribuidor
- US Distribution including PR and to the states of : LA, TN, CA, SC, OR, WA, OK, NV, HI, OH, TX.
- Lote
- Lot Numbers: 15075668, 15075924, 15115426, 15125175, 16025324, 16035493, 16055261.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDADrainer(R) Centesis Catheters: (a) REF 8810, 5F x 9cm (b) REF 8811, 6F x 9cm (c) REF 8812, 5F x 15cm Product27 de junio de 2018
- FDAAbbVie PEG, PEG Kit 15FR, FOR ENTERAL USE ONLY, STERILE, Rx Only.7 de junio de 2017
- FDAThe Acumed Slide-Loc Anatomic Radial Head System is intended for use by surgeons with orthopedic training. The7 de junio de 2017
- FDAGrebSet Micro-Introducer Kit, model 7941, Sterilized with Ethylene Oxide. Product Usage: The GrebSet1 de marzo de 2017
- FDAVascular Solutions Twin-Pass .023" Dual Access Catheter, model 5230. Sterilized with Ethylene Oxide. The26 de octubre de 2016
- FDAVascular Solutions Twin-Pass RX Dual Access Catheter, model 5210. Sterilized with Ethylene Oxide. The26 de octubre de 2016
Preguntas sobre este retiro
¿Dolphin Inflation Device, REF 0185NA. Rx only, Sterilized using Ethylene Oxide. DOLPHIN is a single-use, sterile inflation device used in cardiovascular procedures to pressurize (inflate) and depressurize (deflate) balloon catheters. The device itself consists of two components - the inflation device and large volume syringe (30 cc). The manually operation of the DOLPHIN is achieved by the manipulation of a large handle to drive a piston housed within the body of the device. Careful and controlled inflation is achieved by rotating the handle clockwise. During inflation a unique cam locking mechanism maintains pressure even if the user lets go of the device. Instantaneous deflation, regardless of balloon size, is made possible by the release of the dual locks located on device as well as a large, 30 cc syringe that is sold with each Dolphin Inflation Device. All the while during inflation or deflation pressure is displayed and can be monitored on a large analog gauge mounted on top of the device. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2180-2017 el 7 de junio de 2017. Peligro reportado: The manufacturer of the DOLPHIN inflation device, Perouse Medical, has initiated a recall of the device due to a potential defect in the sterile barrier. Use of affected product could pose a risk of microbial contamination, leading to infection.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter