Target Detachable Coils: TARGET HELICAL NANO 2.5MM X 4CM MODEL Number:M0035452540 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Publicado el 20 de julio de 2016 · Inicio del retiro: 6 de junio de 2016
Resumen
FDA registra el retiro Z-2183-2016 de dispositivos médicos de Clase II, publicado el 20 de julio de 2016 e iniciado el 6 de junio de 2016. Empresa responsable: Stryker Neurovascular; Estado: Finalizado; Cantidad reportada: 45,302 (8,188 in the U.S.) - total, all model numbers; Lote: MODEL Number:M0035452540; Lot Number/Expiration Date: 18440351 30-Jun-18 18539018 31-Aug-18 18439805 30-Jun-18 18519562 31-Jul-18 18468754 30-Jun-18 18922401 31-Mar-19 19060307 31-May-19 18970593 30-Apr-19 18898116 28-Feb-19 19034354 31-May-19 18875384 28-Feb-19. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Stryker Neurovascular
- Distribuidor
- Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.
- Lote
- MODEL Number:M0035452540; Lot Number/Expiration Date: 18440351 30-Jun-18 18539018 31-Aug-18 18439805 30-Jun-18 18519562 31-Jul-18 18468754 30-Jun-18 18922401 31-Mar-19 19060307 31-May-19 18970593 30-Apr-19 18898116 28-Feb-19 19034354 31-May-19 18875384 28-Feb-19.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAINZONE DETACHMENT SYSTEM, REF: M003451009501 de julio de 2026
- FDAThe Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaque16 de octubre de 2024
- FDAThe Synchro SELECT Guidewires are a steerable guidewire family with a shapeable tip and are available in12 de junio de 2024
- FDAThe Synchro Neuro Guidewires are a steerable guidewire family with a shapeable tip. The outside diameter of12 de junio de 2024
- FDATarget Detachable Coils: TARGET HELICAL NANO 2MM X 4CM MODEL Number:M0035452040 Neurology: Target Detachable20 de julio de 2016
- FDATarget Detachable Coils: TARGET HELICAL NANO 3MM X 4CM MODEL Number:M0035453040 Neurology: Target Detachable20 de julio de 2016
Preguntas sobre este retiro
¿Target Detachable Coils: TARGET HELICAL NANO 2.5MM X 4CM MODEL Number:M0035452540 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2183-2016 el 20 de julio de 2016. Peligro reportado: Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter